            both the excitation light for fluorescence and the light used for absorbance measurements. Light of the
            selected wavelength passes through a third half mirror, and is focused onto the sample cell. The light
            reflected at right angles from the third half mirror is focused onto another photoelectric cell. The output
            from this second photoelectric cell provides a reference for fluorescence measurements. The fluorescent
            light emitted at right angles to the excitation light is focused onto a monochromator and the dispersed
            light measured by a diode array. Light reflected at right angles from the second half mirror is focused
            through an optical fiber into the sample cell, and the transmitted light passes to a second
            monochromator, and dispersed onto another diode array.



































                                                         Figure 7.22
                                              The Spectroscopic System that Measures
                                          UV Absorption, Fluorescence and Luminescence
                              Reprinted with permission from M. C. Yappert, M. W. Schuyler and J. D. Ingle,
                                Jr. Anal. Chem. 61(1989)593, Copyright 1989 American Chemical Society