Page 127 - Vibrational Spectroscopic Imaging for Biomedical Applications
P. 127

Evanescent W ave Imaging   103


        analyze materials with little or no sample preparation and the ability
        to analyze highly absorbing materials. With ATR, the only sample
        requirement is that it must be placed into optical (intimate) contact
        with the IRE. In addition, the limited depth to which the evanescent
        beam penetrated the sample meant that spectra of strongly absorbing
        materials could be obtained without total absorption of the infrared
        radiation at a particular wavelength. Microscopic ATR methods did
        not become available until 1991 when Harrick developed the Split-pea
                         19
        infrared microscope  and Spectra-Tech independently developed a
                                                      20
        specialized ATR objective for their IRPLAN microscope.  The Split-pea
        employed a germanium or silicon hemisphere with a beveled tip to
        improve contact with the sample and the appearance of the IRE led to
        the name of the device. The Spectra-Tech ATR objective employed a
        zinc selenide IRE and later a diamond IRE so that the user could
        observe the sample in white light prior to conducting an ATR analysis.
        Perkin Elmer later developed a dropdown accessory for their micro-
        scope, which was based on a germanium hemisphere possessing a
        beveled tip. The user simply aligns the sample in white light viewing
        mode and then lowers the IRE onto the sample for subsequent infrared
        analysis. An added benefit of these devices stems from the fact that the
        pressure applied to a given sample is the force divided by the area.
        Since the contact area is on the order of 100 to 200 μm for each device,
        the pressure and therefore the contact of the IRE with the sample
        increased tremendously as compared to a macro sampling accessory.
        At that time, the major focus of the devices was on the ability to collect
        infrared spectra from intractable samples and not necessarily the
        improvement in spatial resolution.
            The first reports to study the improved spatial resolution of an
        infrared ATR measurement using a germanium IRE was that by
        Nakano and Kawata. 21,22  The authors built a specialized evanescent
        wave microscope that incorporated a confocal aperture for both the
        source and primary image of the sample to spatially isolate the
        sample of interest (Fig. 4.2). The hemisphere with attached sample
        was translated beneath the microscope using a piezoelectrically
        controlled stage. As shown in Fig. 4.2, when the hemisphere is on
        axis, rays enter the hemisphere normal to its surface and, as such,
        are focused at the center of the plano surface. Moving the hemi-
        sphere off-axis to either side, the rays enter at a slight angle, are
        refracted, and come to a focus at off-axis positions, thereby allowing
        different sample points to be interrogated. The authors demonstrated
        an improvement in spatial resolution equal to the refractive index of
        germanium (4×) and the ability to scan over an area of approximately
        100 μm. The limited scan length was the result of spherical aberrations
        introduced by scanning the hemisphere off-axis. In 1995, Esaki et al.
        employed a chevron-shaped internal reflection element (Fig. 4.2) on a
                              23
        conventional microscope.  Esaki et al. demonstrated the ability to
        obtain ATR maps as large as 400 × 400 μm. However, since a hemisphere
   122   123   124   125   126   127   128   129   130   131   132