Page 130 - Vibrational Spectroscopic Imaging for Biomedical Applications
P. 130

106    Cha pte r  F o u r


             FPA









             0.87× Imaging
                Mirror


             Infrared Source




             15× Objective








             Ge hemisphere

        FIGURE 4.3  ATR Imaging using an on-axis confi guration with an array detector
        and the amide I image of a human erythrocyte. [A. J. Sommer, L. G. Tisinger,
        C. Marcott, and G. M. Story, Applied Spectroscopy, Vol. 55, No. 3, page 253,
        fi gure 1 (Society for Applied Spectroscopy, Frederick, Md., 2001).]
        (Permission granted.)


        using an MCT-array-based infrared microscope coupled to a step
        scan interferometer. As depicted in Fig. 4.3, the germanium hemisphere
        was held on-axis. In this mode, the sample was globally illuminated
        and the pixel size of the detector served to spatially isolate a given
        point on the sample.  A comparison of several different  sampling
        modes was conducted, which demonstrated that the ATR mode
        using the array-yielded near-diffraction-limited performance. The
        theoretical and measured spatial resolutions differed only by a
        factor of 1.3. Taking into account the magnification of the system
        from sample to detector, an area approximately 75 × 75 μm could be
        imaged in a matter of minutes. Sommer and co-workers demonstrated
        the capabilities of the system by measuring the surface image of a sin-
        gle human red blood cell. They further showed that the signal
        sensed by one pixel arose from a sample volume of 11 femto-liters.
        This volume relates to a mass detection limit of 13 femto-grams,
                                     3
        assuming a density of 1.2 g/cm . The results of this work were
   125   126   127   128   129   130   131   132   133   134   135