Page 133 - Vibrational Spectroscopic Imaging for Biomedical Applications
P. 133
Evanescent W ave Imaging 109
Refractive Critical Spatial
Material Index Angle θ Resolution ∗
c
Air 1.0 Infinity 4.1λ
Zinc selenide (ZnSe) 2.4 39 1.7λ
Diamond 2.4 39 1.7λ
Si 3.4 26 1.2λ
Ge 4.0 22 λ
∗ Using an aperture splitting beam splitter.
TABLE 4.1 Critical Angle Required for Various IRE Materials
employed, the light would transmit through the IRE/sample inter-
face and no ATR spectrum would be observed. For Si, a portion of the
light would be transmitted and a portion would be internally
reflected. Germanium has the highest refractive index and provides
the best improvement in spatial resolution of all materials. In prac-
tice, germanium is the preferred material but the material is not trans-
parent to visible light, which prevents direct viewing of the sample.
Spectra-Tech, SENSIR (Smith’s Detection), and Varian have opted to
design specialized objectives based on either diamond or a combina-
tion of zinc selenide and diamond. These objectives allow the user to
view the sample and diamond is almost indestructible as an IRE
material.
In general, two approaches have been taken in ATR imaging, on-
axis imaging and off-axis imaging. With on-axis imaging the hemi-
sphere/sample composite is centered at the microscope’s focus and
the hemisphere/sample is illuminated globally. Radiation that is
internally reflected is then imaged onto a two-dimensional array
detector. The detector size defines the sample area that can be imaged
and the pixel size defines the spatial element on the sample, commonly
referred to as the pixel resolution. For example, Sommer employed a
31
64 × 64 MCT array possessing a pixel size of 64 × 64 μm. Based on
these values and the magnification from the detector to the sample,
the area that could be imaged was 76 × 76 μm with a pixel resolution
of 1.2 × 1.2 μm. To increase both the area imaged and the pixel resolution,
one can employ a larger array with smaller pixels.
For off-axis imaging, the hemisphere/sample composite is initially
centered at the microscope’s focus and then imaging is conducted by
moving the composite off-axis as discussed earlier (Fig. 4.2). Lewis
and Sommer demonstrated that for a germanium hemisphere a
1- μm stage displacement will displace the beam in the hemisphere
26
by 0.3 μm. This off-axis mode is employed with either a single
point detector or a linear array detector. The pixel resolution at the