Page 110 - Characterization and Properties of Petroleum Fractions - M.R. Riazi
P. 110

T1: IML
                           QC: IML/FFX
              P2: IML/FFX
  P1: IML/FFX
                        AT029-Manual-v7.cls
            AT029-Manual
                                                        14:23
                                           June 22, 2007
  AT029-03
         90 CHARACTERIZATION AND PROPERTIES OF PETROLEUM FRACTIONS
                                                              gas such as helium or nitrogen. The stationary phase is either
                  250
                                                              solid or liquid. A component that is more strongly attracted to

                  200          ASTM D86                       the mobile phase than to the stationary phase is swept along
                                                              with the mobile phase more rapidly than a component that is
                               SD (ASTM D2887)                more strongly attracted to the stationary phase. The mobile
                 Temperature, °C  150                         phase can be a liquid phase as well; in this case the chro-
                                                              matography method is called liquid chromatography (LC).
                                                                The basic elements of a GC are a cylinder of carrier gas,
                  100
                                                              flow controller and pressure regulator, sample injector, col-
                                                              umn, detector, recorder, and thermostats for cylinder, col-
                   50                                         umn, and detector. The sample after injection enters a heated
                                                              oven where it enters the GC column (stationary phase). The
                   0                                          eluted components by the carrier gas called effluents enter
                                                              a detector where the concentration of each component may
                     0    20    40    60    80    100
                                                              be determined. The presence of a component in the carrier
                               Percent Vaporized
                                                              gas leaving the column is continuously monitored through
                 FIG. 3.4—Simulated and ASTM D 86 distilla-   an electric signal, which eventually is fed to a recorder for a
               tion curves for a petroleum fraction. (The per-  visual readout.
               cent is in vol% for ASTM D 86 and is in wt% for  There are two types of columns, packed or capillary
               ASTM D 2887.)                                  columns, and two types of detectors, flame ionization detec-
                                                              tor or thermal conductivity detector. Packed columns have
         reproducible and can represent the boiling range of a  inner diameters of 5–8 mm and length of 1–5 m. Column
         petroleum mixture without any ambiguity. This method is ap-  and detector types depend on the nature of samples being
         plicable to petroleum fractions with a FBP of less than 538 C  analyzed by the GC. The capillary columns are equivalent to
                                                         ◦
         (1000 F) and a boiling range of greater than 55 C (100 F) and  hundreds of theoretical equilibrium stages and can be used in
                                               ◦
             ◦
                                                     ◦
         having a vapor pressure sufficiently low to permit sampling at  preference to packed columns. The inner diameter of capil-
         ambient temperature. The ASTM D 2887 method is not appli-  lary columns is about 0.25–0.53 mm and their length is about
         cable to gasoline samples and the ASTM D 3710 test method  10–150 m. The stationary phase is coated on the inside wall
         is recommended for such fractions. Distillation curves by SD  of columns. The flame ionization detector (FID) is highly sen-
         are presented in terms of boiling point versus wt% of mixture  sitive to all organic compounds (10 −12  g) but is not sensitive
         vaporized because as described below in gas chromatogra-  to inorganic compounds and gases such as H 2 O, CO 2 ,N 2 ,
         phy composition is measured in terms of wt% or weight frac-  and O 2 . The FID response is almost proportional to the mass
         tion. Simulated distillation curves represent boiling points of  concentration of the ionized compound. Hydrogen of high
         compounds in a petroleum mixture at atmospheric pressure;  purity is used as the fuel for the FID. The thermal conduc-
         however, as will be shown later SD curves are very close to ac-  tivity detector (TCD) is sensitive to almost all the compounds
         tual boiling points shown by TBP curves. But these two types  but its sensitivity is less than that of FID. TCD is often used
         of distillation data are not identical and conversion methods  for analysis of hydrocarbon gas mixtures containing nonhy-
         should be used to convert SD to TBP curves. In comparison  drocarbon gases. The retention time is the amount of time re-
         with ASTM D 86, the IBP from a SD curve of a petroleum mix-  quired for a given component spent inside the column from
         ture is less than IBP from ASTM D 86 curve, while the FBP  its entrance until its emergence from the column in the efflu-
         from SD curve is higher than the FBP from ASTM D 86 of the  ent. Each component has a certain retention time depending
         same mixture (see Fig. 3.4). This is the same trend as that of  on the structure of compound, type of column and station-  --`,```,`,``````,`,````,```,,-`-`,,`,,`,`,,`---
         TBP curves in comparison with ASTM curves as was shown  ary phase, flow rate of mobile phase, length, and tempera-
         in Fig. 3.3. A typical SD curve for a gas oil sample is shown  ture of column. More volatile compounds with lower boil-
         in Fig. 3.4. Note that in this figure the percent vaporized for  ing points have lower retention times. Detector response is
         ASTM D 2887 (SD) is in wt% while for the ASTM D 86 curve  measured in millivolts by electric devices. The written record
         is in vol%.                                          obtained from a chromatographic analysis is called a chro-
          The gas chromatography technique is a separation method  matograph. Usually the time is the abscissa (x axis) and mV
         based on the volatility of the compounds in a mixture. The GC  is the ordinate (y axis). A typical chromatograph obtained
         is used for both generation of distillation curves as well as to  to analyze a naphtha sample from a Kuwaiti crude is shown
         determine the composition of hydrocarbon gas or liquid mix-  in Fig. 3.5. Each peak corresponds to a specific compound.
         tures, as will be discussed later in this chapter. For this reason  Qualitative analysis with GC is done by comparing retention
         in this part we discuss the basic function of chromatography  times of sample components with retention times of reference
         techniques and elements of GC. In an analysis of a mixture by  compounds (standard sample) developed under identical ex-
         a GC, the mixture is separated into its individual compounds  perimental conditions. With proper flow rate and tempera-
         according to the relative attraction of the components for  ture, the retention time can be reproduced within 1%. Ev-
         a stationary and a mobile phase. Recent advances in chro-  ery component has only one retention time; however, compo-
         matography make it possible to identify and separate com-  nents having the same boiling point or volatility but different
         pounds with boiling points up to 750 C (1380 F). A small fluid  molecular structure cannot be identified through GC analysis.
                                      ◦
                                             ◦
         sample (few microliters for liquid and 5 mL for gas samples)  In Fig. 3.5, compounds with higher retention time (x coordi-
         is injected by a needle injector into a heated zone in which  nate) have higher boiling points and the actual boiling point
         the sample is vaporized and carried by a high-purity carrier  or the compound can be determined by comparing the peak












   Copyright ASTM International
   Provided by IHS Markit under license with ASTM             Licensee=International Dealers Demo/2222333001, User=Anggiansah, Erick
   No reproduction or networking permitted without license from IHS  Not for Resale, 08/26/2021 21:56:35 MDT
   105   106   107   108   109   110   111   112   113   114   115