Page 241 - Advances in Forensic Applications of Mass Spectrometry - Jehuda Yinon
P. 241

1522_book.fm  Page 214  Thursday, November 13, 2003  9:58 AM









                             Chromatogram Plots
                             File: I:\data0999\15704 ms2.ms
                             Sample: 10.0 MIN SAMPLING TIME, 5 MINDESORB, GB      16/9/99 10:37 AM
                             Sample Notes: RN0391-01, 99-0817-F #1, STAIR TREADS - GC/MS/MS - GB
                             Operator: D.SUTHERLAND                  Scan Range: 1-1034 Time Range: 0.02 - 15.00 min.
                                  Slecl 1               7.099 m In. Scan: 343 Chan: 1 Ion: NARIC: 12512
                                  B P 131 (3894-100%) 15704 ms2.ms  131
                              100%
                              75%                            119
                              50%
                                                        105
                              25%                  91
                                       57     75  84                    147   165   181  194  202
                               0%
                                  Slecl 2                8.126 m In. Scan: 385 Chan: 1 Ion: NARIC: 10869
                                  B P 131 (2651-100%) 15704 m : 2.ms  131
                              100%
                              75%                       105
                              50%
                                                 84          119
                              25%                  91                  146
                               0%    50  57  66                             158  174  184  192  201 208
                             D Iff  Slecl 1 Specl 2     7.099 m In. Scan: 343 Chan: 1 Ion: MARIC: 12512
                              100%  Normalized          8.126 m In. Scan: 393 Chan: 1 Ion: MARIC: 10869
                              50%
                               0%
                              −50%
                             −100%
                                     50       75       100       125      150      175      200 m/z
                             kCounts                                        15704 ms 2.ms Ions: 131 all
                                3
                                2
                                1
                             Counts                                         15704 ms 2.ms Ions: 146 all
                               700
                               600
                               500
                               400
                               300
                               200
                               100
                                         6             7            8             9        minutes
                                                      2                               3
                                         289           338          387           433       Scans
                             Figure 5.18  Comparison of GC/MS/MS spectra for a third indane isomer: ion
                             ratios different.
                            Table 5.3 GC/MS/MS — Target Class Approach
                              Center of  Window    CID  Example parent
                              parent ion  Range   Energy  or daughter
                             window (m/z)  (+/– m/z)  (V)  ion (m/z)       Target chemical classes
                                 71         3       8   71, 69        Alkanes, alkylcyclohexanes
                                106         5      35   91, 105       Alkylbenzenes
                                118         5      35   117, 118      Indanes, alkylbenzenes
                                133         5      35   119, 133      Alkylnaphthalenes, alkylbenzenes
                                145         7      40   141, 142, 145  Alkylbenzenes, indanes
                                158         5      35   154, 156      Alkylnaphthalenes, alkylbiphenyls
                                169         5      45   155, 170, 168  Indanes, alkylbiphenyls
                                183         5      40   156, 171      Alkylnaphthalenes, alkylanthracenes
                             that an ASTM method does not yet exist for GC/MS/MS as investigators and
                             reviewers are not familiar with this method, and that there needs to be an
                             adequate time of comparison of the method with the current benchmark of
                             GC/MS.
                                A single GC/MS/MS method can be used to analyze fire debris as is being
                             discussed in a sub-committee of ASTM E30.01 that is charged with the review

                             © 2004 by CRC Press LLC
   236   237   238   239   240   241   242   243   244   245   246