290   Control theory in biomedical engineering


          patients with NPC. The NasoXplorer aims to monitor and increase the fre-
          quency of observations of NPC while reducing the burden on healthcare
          professionals. We focus on the mechanics of the NasoXplorer and will
          not discuss further considerations for camera models and electronic transfer
          or quality of captured images. For the final prototype, we used optical fibers
          to test the viewing angle of the NasoXplorer. Although the NasoXplorer has
          only 2 degrees of freedom, it has a wide viewing range of up to 90degrees.


          2.1 Device specifications from anatomical considerations
          Besides user needs and intended use of a device, other design considerations
          had to be made to consider the different anatomical variations of different
          people. We observed that several characteristics played a part in anatomical
          differences.


          2.1.1 Anatomical variations in shape
          The NasoXplorer aims at minimizing end-user discomfort, especially for a
          self-administered device. As the distal segment of the device will seek to
          work as a probe, a flexible and soft material was deemed most suitable for
          this function. The flexibility of the NasoXplorer allows the camera to have
          a broader optical viewing range, while the softness of the device prevents
          significant user discomfort. By minimizing the size of the NasoXplorer,
          patients with different nose shapes are easily be able to navigate the naso-
          pharynx region. Hence, the functionality of NasoXplorer is maximized in
          the observation of the nasal cavity and nasopharynx region.


          2.1.2 Anatomical variations in size
          Anatomical differences exist between individuals. For the NasoXplorer to
          cater to different individuals, anthropometric data had to be studied. As
          the NasoXplorer is designed to be self-administered, the reduction of
          end-user discomfort was of importance. This meant that we had to
          cater to all patients regardless of anatomical differences. The rigidity of
          the NasoXplorer provides a fuss-free, easy-to-use solution for patients.
          A softer, flexible tip was employed to allow monitoring of the more
          in-depth part in the nasopharynx region. However, our device may not
          be suitable for viewing the deeper regions in the nasopharynx region. This
          can be improved in the future by using a more extended tip. However,
          adjustments may have to be made to the length of the spring as a more
   311   312   313   314   315   316   317   318   319   320   321