Page 318 - Control Theory in Biomedical Engineering
P. 318

292   Control theory in biomedical engineering


                        Anatomical         Design
          Parameters    restrictions       specifications  Comments
          Diameter      Should ideally be  3.92mm        Parameter met with
            (outer)       less than 10mm                   2mm of space
                          Data collected                   between device and
                          was based on                     lumen on each end
                          Caucasians,
                          should be
                          adjusted to 8mm
          Diameter                         3.9mm         The diameter of the
            (inner)                                        camera is 3.9mm,
                                                           too large for the
                                                           aluminum tube.
                                                           Further
                                                           improvements can
                                                           be made to use a
                                                           smaller version of
                                                           the IntroSpicio
                                                           camera instead
          Insertion     Minimum of         8.5mm         Parameter not met,
            length        100mm                            but still sufficient to
            Length of                                      capture more than
            workspace                                      80% of the nasal
            in region                                      cavity
            (sagittal)                                     Improvements can
                                                           be made to lengthen
                                                           the flexible tip
          Bending                          145           Parameters met
            angle
            (degrees)
          Control                                        Further improvements
            system                                         will be made to
                                                           contain the control
                                                           system and improve
                                                           the portability of the
                                                           device



          2.2 Overall design

          Our device in Fig. 1 encompasses the traits of a flexible nasopharyngoscope,
          which includes a small diameter of 3mm and a high deflection angle of
          145degrees (Table 1). Both the diameter and deflection angles are compa-
          rable to those that are commercially available.