Continuum NasoXplorer manipulator with shape memory actuators  293


























              Fig. 1 The prototype, optical image (top), CAD model (middle) and illustrations
              (bottom).


              Table 1 NasoXplorer specifications.
              NasoXplorer
              Diameter (tip)/mm                              3
              Diameter (proximal end)/mm                     8.4
              Angulation range (deflection)/degrees          145
              Working length (active)/cm                     2
              Working length (passive)/cm                    9.8



              2.3 Design components and design rationale
              2.3.1 Optical zooming segment and camera
              The camera we used is the IntroSpicio 110 micro CCD camera. It has a
              diameter of 3.9mm and is threaded through the TangoBlack flexible seg-
              ment, as well as the sole SMA spring in the NasoXplorer (Fig. 2). The distal
              end of the NasoXplorer is the flexible tip, bending up to an angle of
               100degrees. This is supported by the User and Technical Specification
              Report. The NasoXplorer works using an SMA spring that extends and
              contracts with the application of a small current. Fig. 3 shows part of the
              NasoXplorer when relaxed and contracted, respectively. The spring used
              in the NasoXplorer is 0.1in. in diameter and 0.6in. in length. Further details
              are provided in Fig. 3. When the NasoXplorer enters the nasal inlet of the