          Fig. 17 Illustration of the intended usage of the prototype in a skeleton model. (Inset
          bottom) Set-up for insertion test (side view); inset top-left: set-up for insertion test
          (top view).
          illumination was used to confirm if the NasoXplorer could capture images
          where the points were if it had a camera inside.
             Using this set-up, the NasoXplorer successfully pointed at all the points
          inside the skull. Hence, it can capture all parts of the nasopharyngeal region.



          Summary of design verification tests and associated acceptance criteria.
                           Related
                           methods
          Testing subjects  of testing  Acceptance criteria  Results after testing
          Breakage of      ASTM       No breakage of the  Due to the
            flexible tip     F2942-     tip. Breakage       reinforced tip, the
            during           13         could potentially   flexible tip was
            endoscopy                   cause harm as it    much sturdier, and
            (due to cyclic              may cause the       no breakage was
            bending)                    device to be in     observed
                                        contact with
                                        sensitive tissue
          Wire breakage    ASTM       It does not break   No breakage was
            during           A931-      after repeating     observed
            endoscopy        18         cycles
                             ASTM
                             B1-13
                             (2018)


          4 Design review

          We conducted a Design Review to study design results, serve as feedback
          control, and assess the progress of the project. The NasoXplorer provides