Page 331 - Control Theory in Biomedical Engineering
P. 331
304 Control theory in biomedical engineering
Control setup
Power supply
Prototype in
intended
application on
model
Fig. 17 Illustration of the intended usage of the prototype in a skeleton model. (Inset
bottom) Set-up for insertion test (side view); inset top-left: set-up for insertion test
(top view).
illumination was used to confirm if the NasoXplorer could capture images
where the points were if it had a camera inside.
Using this set-up, the NasoXplorer successfully pointed at all the points
inside the skull. Hence, it can capture all parts of the nasopharyngeal region.
Summary of design verification tests and associated acceptance criteria.
Related
methods
Testing subjects of testing Acceptance criteria Results after testing
Breakage of ASTM No breakage of the Due to the
flexible tip F2942- tip. Breakage reinforced tip, the
during 13 could potentially flexible tip was
endoscopy cause harm as it much sturdier, and
(due to cyclic may cause the no breakage was
bending) device to be in observed
contact with
sensitive tissue
Wire breakage ASTM It does not break No breakage was
during A931- after repeating observed
endoscopy 18 cycles
ASTM
B1-13
(2018)
4 Design review
We conducted a Design Review to study design results, serve as feedback
control, and assess the progress of the project. The NasoXplorer provides