Page 245 - Academic Press Encyclopedia of Physical Science and Technology 3rd Analytical Chemistry
P. 245

P1: GRB/GWT  P2: GPJ/GAX  QC: GAE/FYD Final Pages
 Encyclopedia of Physical Science and Technology  EN008M-395  June 29, 2001  15:52






               970                                                                           Magnetic Resonance in Medicine


                                          2
               for  water  is  D = 3.7 × 10 −5   cm /sec.  During  a  time  t  II.  NUCLEAR MAGNETIC
               the  random-walk  motion  associated  with  self-diffusion  RESONANCE SCANNERS
               will give each molecule an average displacement equal  AS MEDICAL INSTRUMENTS
                 √
               to  Dt. For t = 10 msec this gives an average displace-
               ment of 6 µm and for t = 100 msec, 19 µm. These dis-  A.  System Considerations
               tances are comparable with average cell sizes. This shows
                                                                 The overall operation of a NMR scanner is controlled by
               that during a typical T 2  interval a given water molecule
                                                                 a computer (Fig. 3). It provides the pulse-timing infor-
               has time to encounter much of the intracellular and local
                                                                 mation for the gradient and radio-frequency transmitter
               extracellular environment, and therefore to average out
                                                                 coils. It also switches on the preamplifier and the receiver
               the inhomogeneities in local field implied by the cellular
                                                                 circuitry during the time when data is being acquired from
               heterogeneity.
                                                                 the nuclear spins. The data are acquired as free induction
                                                                 decay or, more commonly, spin echo signals. Extensive
                                                                 mathematical  calculations,  usually  a  two-dimensional
               G.  Chemical Shift
                                                                 Fourier transformation, are required to convert the FID
               As discussed, the electrons orbiting each atom respond in  or spin echo data into an image and the calculations are
               a diamagnetic fashion when an external field is applied.  done  by  the  computer.  For  permanent  magnet  systems
               As a result, the total magnetic field at the nucleus of an  no magnet power supply is required; for superconducting
               atom is some what less than the externally applied field.  magnet systems the magnet power supply is needed only
               The exact value of this reduction depends on the details  initially when the magnet is being energized.
               of the electron orbits in the given molecule and will be  The NMR signal is sufficiently weak that interfering
               different in different molecules. Consequently, in a given  electrical  noise  is  a  major  consideration.  To  minimize
               external field and for a given type of nucleus, the nuclear  outside  electrical  interference,  some  form  of  electrical
               resonant frequency will vary from one type of molecular  screening is normally placed around the scanner or the
               environment to another. This change in frequency δf  is  scan room, and electrical filters are used on those circuits
               called the chemical shift. It is usually small, and δf /f 0  that could transmit outside noise to the receiver. Figure 4
               can be specified in parts per million. These small shifts  shows a patient within a screened room being positioned
               are easily measured, however, and in MR spectroscopy  for scanning. In Fig. 5 the patient is seen within the magnet
               they are the basis for discriminating one molecular species  bore in the location that permits scanning to be done.
               from another. In MRI the chemical shift leads to an image  The strength of NMR signal generated by the precess-
               artifact that slightly confuses the boundaries between fat  ing spins is proportional to the degree of alignment of
               and water-containing tissues. Protons in water and fat have  the spins [M 0 in Eq. (2)] and to their rate of precession
               resonant frequencies differing by 3.5 ppm.        (γB 0 ). Since both those quantities are proportional to field




























                                  FIGURE 3  Block diagram of an NMR scanner. (Courtesy of GE Medical Systems.)
   240   241   242   243   244   245   246   247   248   249   250