Page 249 - Academic Press Encyclopedia of Physical Science and Technology 3rd Analytical Chemistry
P. 249

P1: GRB/GWT  P2: GPJ/GAX  QC: GAE/FYD Final Pages
 Encyclopedia of Physical Science and Technology  EN008M-395  June 29, 2001  15:52






               974                                                                           Magnetic Resonance in Medicine


               regulations. During the 1980s several manufacturers suc-  useofstaticfieldsupto4.0Tand,underappropriatelycon-
               cessfully sought FDA approval for their scanners.  trolled conditions, an average heat input to the patient’s
                 As additional experience was gained during the 1990s,  body (specific absorption rate or SAR) of 4.0 W/kg. The rf
               FDA approval was granted for a variety of scanner en-  heating that occurs under standard imaging conditions is
               hancements, such as the use of higher field strengths and  comparable to that resulting from normal metabolic activ-
               a wider variety of RF and gradient coils.         ity and is unnoticeable to the patient. Likewise, the possi-
                 NMR scanners place the patient in an environment that  ble effects of dB/dt on nerve excitation and the magneto-
               is quite unlike that of any other medical instrument. Ini-  hydrodynamic interaction between flowing blood and the
               tially, there were several areas of concern that, with expe-  static magnetic field appear to require conditions far be-
               rience, have become better understood and appear not to  yond those used in modern scanners. At the present time
               represent a danger to patients. The area of greatest contin-  these concerns represent hypothetical, rather than actual,
               uing concern is the interaction between the strong, static,  hazards of the scanning process.
               magnetic field and ferromagnetic substances inadvertently
               brought into the region of the scanner. One version of
               this problem arises from the fringing field surrounding  III. IMAGING TECHNIQUES
               the magnet, which can be treacherously strong. Many ob-
               jects common in hospitals (e.g., oxygen bottles, mops,
                                                                 A. Selective Excitation
               fans, and hairpins) contain enough magnetic material that
               they can be drawn into the magnet with great force and  One of the key capabilities of MRI scanners is the ability
               rapidity. Such flying objects are extremely dangerous to  to excite a single, thin slice of spins within the patient. This
               anyone in their path. For this reason, most manufacturers  permits the construction of images that have the character
               and users of the scanners go to great lengths to limit access  of two-dimensional cross-sectional cuts through the pa-
               to the vicinity of the scanner. Permanent magnet systems,  tient’s anatomy. NMR imaging resembles CT scanning in
               and some superconducting systems that have magnetic  producing this type of anatomical image. However, MRI
               shielding around them, have smaller fringing fields and  has a substantial advantage over CT in that it permits the
               are less susceptible to this effect. A second version of this  location of the slice to be chosen electronically by the op-
               problem comes about because some patients have ferro-  erator without moving the patient or any components of
               magnetic substances implanted within their bodies. This  the scanner. In MRI, imaging planes of any orientation
               is usually the result of a prior surgical procedure, such as  may be chosen, and these planes may also be moved elec-
               the clipping of a diseased blood vessel, but in some cases  tronically from side to side, top to bottom, or front to back
               iron fragments (e.g., shrapnel) have become embedded  through the patient’s anatomy.
               in a patient’s tissues during some sort of accident or ex-  The key to selective excitation is carrying out the rf exci-
               plosion. Patients are not always aware of the presence of  tation in the presence of a gradient field. Suppose the static
               these objects. Therefore care has to be taken before scan-  field is uniform across the patient. If a rf pulse is applied at
               ning to exclude those patients with possibly dangerous fer-  the Larmor frequency, ω 0 = γB 0 , for a time long enough
               romagnetic implants. Implanted cardiac pacemakers can  to create a 90 pulse, this will excite spins over a large
                                                                             ◦
               malfunction, or conceivably, be permanently damaged be-  volume of the patient. If, however, the z-gradient coil is
               cause of exposure to strong magnetic fields. Therefore  used to apply an additional static field, B z = G z z, at the
               patients with these devices in place are not normally can-  time of the rf pulse, the resonance condition will be met
               didates for NMR scanning.                         strictly only in the plane z = 0. Spins far from this plane
                 Other areas of initial concern were the effects of the  are well off resonance and essentially will be unaffected
               static field on normal tissue function; the possibility that  by the rf pulse. Spins at z = 0 will be rotated by 90 just
                                                                                                           ◦
               electric fields generated by the rapidly changing gradient  as if no gradient field were present. Spins close to, but not
               fields (i.e., the “dB/dt” effect) could cause nerve stimula-  at, the plane z = 0 will be partially excited. To predict the
               tion or irregularities in the cardiac rhythm; tissue heating  exact behavior of the excitation, as a function of z, near the
               associated with the rf excitation field; and possible effects  origin it is necessary to carry out a solution to the Bloch
               on blood pressure resulting from forces of interaction be-  equations. It turns out that spins in the selected slice, but
               tween blood and the static magnetic field. In the scanners  slightly off the center plane, will also be flipped through
               presently used, all of these effects appear to be readily tol-  approximately 90 but will have phase differences with
                                                                                ◦
               erable, and in most cases, negligible. The FDA continues  those spins at z = 0. This will reduce somewhat the signal
               to receive and evaluate designs for more advanced scan-  generated by the slice. It can be shown that a more perfect
               ners, and the regulatory aspects of scanner safety continue  slice profile will result if, instead of using a rf pulse of
               to evolve. Of particular note are guidelines approving the  constant amplitude, the pulse amplitude is modulated by
   244   245   246   247   248   249   250   251   252   253   254