Page 259 - Academic Press Encyclopedia of Physical Science and Technology 3rd Analytical Chemistry
P. 259

P1: GTM/GRL  P2: GNH Final Pages
 Encyclopedia of Physical Science and Technology  EN009N-406  July 18, 2001  23:32







              Mass Spectrometry                                                                           147

              electrically to 400 C. Automation of a direct probe sample  different from performing the same specification at levels
                            ◦
              introduction system has recently been described (Manura  of 0.01%. With increased resolution, chromatography is
              and Manura, 2000). With the automated system, samples  better able to isolate components present in mixtures at
              can be introduced to the mass spectrometer at the rate of  lower levels. With increased sensitivity, mass spectrome-
              15–20 samples per hour. This is a faster rate of sample  try is better able to identify such components. Finally, with
              analysis than is normally possible than with chromatogra-  increased emphasis on the measurement of analytical in-
              phy coupled with mass spectrometry, in which the column  formation for both screening and regulatory purposes, the
              separation time accounts essentially for all of the time of  demand for trace analyses has increased significantly.
              the analysis.
                                                                  a. Gas chromatography. The combination of gas
                                                                chromatography with mass spectrometry (GC/MS), itself
                2. Chromatographic Columns
                                                                the topic of an excellent text by McFadden in 1973, is
              The ability of mass spectrometry to identify a sample com-  now realized within an integrated, low-cost, widely avail-
              pound is maximized when the sample is pure; the combi-  able analytical instrument. There is no longer any need to
              nation of mass spectrometry with a chromatographic inlet  review the history of development of GC/MS instrumen-
              system has therefore become a mainstay of instrumental  tation, nor the transition from packed-column to capillary
              analysis. Chromatography separates sample components  column GC. Instead, it is relevant to consider the simple
              in time, passes them into the mass spectrometer, and each  physical interface between the methods, and then con-
              is then characterized via measurement and interpretation  straints on the operation of the methods in the combined
              of its mass spectrum. The first gas chromatography/mass  GC/MS instrument, and finally, to preview the information
              spectrometry (GC/MS) instruments of the middle 1970s  available to the analyst from the GC/MS combination.
              used packed columns, a 5-mm outer diameter, 1-m long  In the modern GC/MS instrument, there is no interface
              glass tube filled with the packing material (silica or di-  per se between the capillary column and the ionization
              atomaceous earth) onto which the stationary phase was  source of the mass spectrometer. The influx of helium
              coated. An enrichment device was needed to separate the  carrier gas is of such low amount that it can be accommo-
              molecules of the sample from the great excess of the he-  dated without difficulty by the vacuum pump connected
              lium carrier gas flow. The resolution achieved with such  to the source, while still maintaining high vacuum in the
              columns was relatively low; peak widths could be tens of  ionization source and the mass analyzer. The capillary col-
              seconds wide, and the occurrence of overlapping peaks in  umn is terminated by direct connection to the ionization
              the separation of a complex mixture common. If we were  source, with all sample from the column passed directly
              still using packed columns for the separation of complex  into the source. Termination of the column at the pres-
              mixtures, mass spectrometry would be limited to analyz-  sure of the mass spectrometer rather than the usual at-
              ing mixtures of 10–20 nonoverlapped components, all ex-  mospheric pressure of many other GC detectors does not
              hibiting the proper thermal stability and volatility for GC  change the retention time significantly, since the pressure
              characterization.                                 drop occurs in only the last few cm of the column. It is
                Clearly, the capabilities of modern chromatographic  important that a bonded, stable stationary phase be used
              techniques have been vastly improved. Packed column  in the capillary column to minimize the amount of column
              GC has been replaced by capillary column GC. Similarly,  bleed.Columnbleediselutionofthestationaryphaseitself
              the large columns of normal-phase liquid chromatogra-  into the ionization source of the mass spectrometer. Sen-
              phy (LC) are replaced by microcolumn reverse phase LC  sitivity is compromised, as is unambiguous identification
              columns. Capillary electrophoresis (CE) is an entirely new  of compounds due to the presence of extraneous ions in
              means of separating small amounts of more complex, and  the mass spectrum formed from the stationary phase. Al-
              charged sample molecules, and has evolved into several  though background subtraction can remove or reduce the
              distinct forms with unique capabilities. Mass spectrom-  contribution of these bleed ions, this requires additional
              etry coupled with different forms of chromatography is  spectral processing time.
              now applied to the analysis of many mixtures, of higher  As eluting peaks from a GC become narrower (as sep-
              complexity, and more disparate sample types.      aration resolution increases), the need for faster scanning
                It is appropriate here to revisit the meaning of “mixture  of the mass analyzer becomes more stringent. Even for
              analysis.” A mixture becomes increasingly complex as  GC peaks only a few seconds wide, at least a few com-
              measured component levels decreases, and as the mixture  plete mass spectra should be recorded so that they can be
              is examined with increasingly sensitive and sophisticated  averaged together to form an approximation of the mass
              methods. Specification of mixture components with levels  spectrum measured with a steady sample concentration in
              greater than 1% provides a snapshot of composition vastly  the ionization source. Modern mass analyzers, including
   254   255   256   257   258   259   260   261   262   263   264