122 Chapter 5
5.3.2.5 Satisfaction benchmarking
Table 5.6: Satisfaction benchmarking table shows the level of satisfaction consumers have
toward each of the prototypes. More “crosses” indicates a higher level of satisfaction, with the
maximum number of “crosses” being five.
Satisfaction benchmarking
Need Customer Importance Olympus Ambu aScope 4 PENTAX
no. need (1 5) MOJO LF TP Broncho Slim FI-7RBS
1 Controlled 1 XXX XXXXX
force
2 Stability 1 XXXX XX XX XX
and rigidity
3 Good 1 Only with the use of XXXXX XXXXX XXXXX
visualization fiberscope bronchoscopy
4 Minimally 1 XXX XXX XXX XXX
invasive
5 Ease of 2 XXX XXX XXXXX XXX
operation
6 Customize 2 XXXX
7 Durability 3 XX XXXXX X XXXXX
8 Affordable 4 XXXXX XXXX XXX XXX
5.3.2.6 Target specification
Table 5.7: Target specifications table serves as a guide for the developers while reviewing and
improving the device. The ideal value is the set of values that we aim to achieve with MOJO to
fulfill user needs. The marginal values are the minimum acceptable values for each specification.
Target specification
Metric Need Importance
no. no. Metric (1 5) Unit Marginal value Ideal value
1 1,2 Automatic 1 Manual/ Automatic
mechanism semi/full
2 2 Stability and 1 N 6 0.01 .0.039 a
stiffness of
the device
3 3 Lumens in 1 No. of 4
the device lumens
4 4 Diameter of 1 mm 5.2
the device
6 6 Bending 1 Degrees Segment 1: 3.9 Segment 1:
angle b Segment 2: 2.8 117
Segment 3: 2.0 Segment 2:
Segment 4: 7.0 137
Segment 3: 90
Segment 4: 90
(Continued)