Page 207 - Fundamentals of Light Microscopy and Electronic Imaging
P. 207

190      FLUORESCENCE MICROSCOPY

                                 • A bright light source such as a mercury or xenon arc lamp is required because only
                                    a narrow band of wavelengths, and consequently a small portion of the illuminator
                                    output, is used to excite fluorochromes in the specimen.
                                 • For efficient high-contrast imaging, both the illuminator and objective lens are
                                    positioned on the same side of the specimen. In this arrangement, the lamp and light
                                    delivery assembly are called an epi-illuminator, and the objective lens functions
                                    both as the condenser, delivering excitatory light to the specimen, and as the objec-
                                    tive lens, collecting fluorescent light and forming an image of the fluorescent object
                                    in the image plane.
                                 • Fluorescence filter sets containing three essential filters (excitation filter, dichroic
                                    mirror, and barrier [or emission] filter) are positioned in the optical path between
                                    the epi-illuminator and the objective. This arrangement is shown in Figure 11-7.
                                 • High-NA, oil immersion objectives made of low-fluorescence glass are used to
                                    maximize light collection and provide the greatest possible resolution and contrast.

                                    Epi-illumination is made possible by the employment of a dichroic mirror, which is
                                mounted together with exciter and barrier filters as a fluorescence filter set in a filter






                                                                           Barrier or
                                                                          emission filter


                                                    Exciter
                                                     filter                     Dichroic mirror
                                          Light
                                         source                               Filter cube









                                                                              Objective


                                                                              Object

                                Figure 11-7
                                Arrangement of filters in a fluorescence filter cube. The diagram shows the orientation of
                                filters in a filter cube in an epi-illuminator for an upright microscope. The excitation beam
                                (dotted line) passes through the exciter and is reflected by the dichroic mirror and directed
                                toward the specimen. The return beam of emitted fluorescence wavelengths (solid line)
                                passes through the dichroic mirror and the emission filter to the eye or camera. Excitation
                                wavelengths reflected at the specimen are reflected by the dichroic mirror back toward the
                                light source. Excitation wavelengths that manage to pass through the dichroic mirror are
                                blocked by the barrier (emission) filter.
   202   203   204   205   206   207   208   209   210   211   212