Page 306 - Fundamentals of Light Microscopy and Electronic Imaging
P. 306

FLAT-FIELD CORRECTION      289

                       response of the eye or camera film and are created using a function with variable expo-
                       nent   such that

                                Displayed value   [(Normalized value 	 Min)/(Max 	 Min)] .

                          The adjustment of   and a figure showing LUTs with various   values were pre-
                       sented in Figure 13-7. With   settings  1, low pixel values are boosted relative to high
                       values and appear a medium gray in the image; this adjustment also reduces the contrast
                       between bright features and the darker background. A   setting of 0.7 approximates the
                       response of the eye, allowing the image to more closely resemble the view we perceive
                       when looking in the microscope. Conversely,   values  1 depress dark and medium
                       gray pixel values and increase the visibility and contrast of bright features. The effect of
                       adjusting   on the image display is shown in Figure 15-3. In Photoshop, the slider used
                       to define the midpoint of gray values performs an identical function. Photoshop gives
                       you the choice of reading off a   value corresponding to the slider’s location or entering
                       a numerical value of your own choosing for  .



                       FLAT-FIELD CORRECTION

                       A single image prior to processing and adjustments is called a raw image. In many
                       cases, raw images are suitable for printing directly, perhaps after first making some
                       minor adjustments to brightness and contrast using the image histogram. However, a
                       raw image contains two significant kinds of noises that make it unsuitable for quantita-
                       tive purposes: First, it contains the bias signal and noise counts that increase pixel val-
                       ues beyond their true photometric values, which can cause significant errors in
                       measuring the amplitudes of object features. Second, a raw image may contain numer-
                       ous artifacts from the camera and microscope optical system (distortions due to detec-
                       tor irregularities, dust and scratches on lens surfaces, uneven illumination) that appear
                       as dark shadows and specks in the image and alter the true pixel values. These artifacts
                       are particularly visible in images with bright, uniform backgrounds (phase contrast,
                       DIC) and in fluorescence images with medium gray or bright backgrounds. In order to
                       restore photometric accuracy and remove the defects, the raw image must be adjusted by
                       an operation known as flat-field correction (Gilliland, 1992; Howell, 1992). Only cor-
                       rected images are suitable for measuring light amplitudes. Although flat-field correction
                       is not required to display or print an image, it is very effective in removing disfiguring
                       faults, so the procedure is widely used for cosmetic purposes as well. As shown in Fig-
                       ure 15-4 and described here, the correction procedure requires three frames:


                        •A raw frame contains the image signal plus signals from optical defects and elec-
                          tronic and thermal noises.
                        •A flat-field frame is the image of a smooth, featureless field without the object, and,
                          if possible, is made at the same focus as the raw frame. Flat frames should be bright
                          and utilize the full dynamic range of the camera in order to minimize the noise in
                          the final corrected image. If both the raw and flat-field frames are dim and noisy,
                          the final processed frame will be very noisy indeed. This situation is particularly
                          relevant to fluorescence imaging. Therefore, the exposure time for a flat-field frame
                          might be longer than that used for producing the raw image itself. To reduce noise
   301   302   303   304   305   306   307   308   309   310   311