Page 116 - Gas Adsorption Equilibria
P. 116

102                                                         Chapter 2


          Temperature gradients during heating up and cooling down of the sample as
          well  as the time of activation should be  chosen carefully as  they may have
          considerable impact on the properties of the activated sorbent material. After
          this pre-treatment  procedure  is  completed the  whole  instrument has  to be
          evacuated. Then sensor gas should be provided to the gas thermometers (1,2).
          After this  the  instrument is  placed within  the thermostat.  When  thermal
          equilibrium is reached valves (7,7A) are closed. At this moment the reading
          of the  difference  pressure  gauge corresponding  to equal  pressures on  both
          sides of the instruments membrane is a stable signal providing the baseline of
          any pressure  signal  measurement and  hence  also  of any calorimetric
          measurement.

             After filling the gas reservoir with an appropriate amount of sorptive gas
          the adsorption process can be started by opening valve (V5). The sorptive gas
          is flowing from the reservoir via capillary (3) into the (evacuated) sorption
          vessel. During this process it is important to make sure that the velocity of the
          gas flow is small enough to allow thermal equilibration of the gas temperature
          to thermostat temperature (T*). Otherwise it would become difficult to set up
          a concise  energy  balance of the  adsorption process  as  the  enthalpy of the
          incoming sorptive gas would not be known exactly. As soon as the sorptive
          gas  reaches the sorbent  material prepared inside  the  adsorption  vessel
          adsorption occurs and the heat of adsorption is released to the sorptive gas and
          then transferred via the wall of the adsorption vessel to the sensor gas inside
          the  sensor  gas vessel, cp.  Fig.  1. As  the  temperature of the  sensor gas  is
          increased due  to this  heat  transfer so  is its pressure   compared to  the
          pressure  in the reference vessel   Hence there is a pressure difference
                              which is monitored and  registered by  the difference
          pressure manometer (P3).  Additionally,  pressures at  manometers (P1) and
          (P2) also are monitored allowing to determine the amount of gas adsorbed i. e.
          to evaluate the adsorption isotherm.

             A numerical example for a pressure difference signal recorded during a
          calibration process of an SGC is given in Fig.  2.11. It shows an increase in its
          first part  reaching a  maximum  value after  which it  again approaches the
          baseline i.  e.  its initial  value indicating  thermal equilibrium  in  the overall
          system.  After this,  sorptive gas again can be added to the adsorption vessel
          increasing thus the inside pressure (p)  and starting a new adsorption process
          which finally will  lead to  a  new  adsorption equilibrium  at  some  higher
          pressure
   111   112   113   114   115   116   117   118   119   120   121