Page 298 - gas transport in porous media
P. 298

296
                           flow in a Venturi tube is:
                                                                                        Rossabi
                                                      πd 2      2 p
                                                 Q =         
                           (17.4)
                                                       4   ρ 1 − (d/D) 4
                                                         3
                           whereQ isthevolumetricflowrate(m /sec),  pisthedifferenceinpressuremeasured
                                                                                       3
                           in the pipe and the pipe constriction (Pa), ρ is the density of the fluid (kg/m ), and d
                           and D are the diameters of the constriction and the main pipe respectively (m).
                             The expression for fluid velocity as measured by a Pitot tube is:

                                                             2 p
                                                        ν =                              (17.5)
                                                               ρ
                           wherev isthevelocityofthefluid(m/sec), and pinthiscaseisthedifferencebetween
                           absolute and static pressure measured at the stagnation point (Pa). In general, losses
                           increase from Pitot and Prandtl tubes to Venturi meters to nozzles to orifice plates.

                           17.4 THERMAL FLOW SENSORS

                           Because of the prevalence of the use of thermal sensors for gas flow, a convenient way
                           to differentiate flow sensors is in terms of thermal and non-thermal mechanisms. The
                           theoretical basis for thermal flow sensors arose from heat conduction work conducted
                           in the early twentieth century. In particular, the work of King (1914) was seminal to
                           the development of in-line and insertion mass flow meters, two of the most popular
                           process flow meters. King developed the expressions for the conduction of heat
                           from radially symmetric surfaces in a constant flow field that are essentially used
                           today.
                             Thermal anemometers use the heat transfer properties of the fluid they are mea-
                           suring to calculate a mass flux rate rather than a volumetric flow rate. The size
                           and number of gas molecules that hit the heated element determine the amount of
                           energy removed from the element. The heated sensor element employed can be a
                           wire, a plate or film, a sphere or another shape that is able to sample the flow yet
                           provide minimal impedance or change to the flow. These devices are deployed in
                           the flow path and are usually referred to as insertion mass flow meters. Their small
                           size also permits easy access to flow fields. Some thermal flow meters use a section
                           of the flow conduit as the measurement device (in-line meters) either as the heated
                           sensor itself or in conjunction with embedded heated elements (Viswanathan et al.,
                           2002).
                             Thermal sensors measure either the amount of energy (usually as electric power)
                           required to maintain a heating element at a constant temperature in the mass flow path,
                           or they measure the temperature of the element for a particular applied energy (electric
                           power) in the mass flow path (Baker and Gimson, 2001).Although the thermal sensors
                           actually measure the mass of fluid interacting with the heated element, they are often
                           calibrated to the velocity of a particular gas compound flowing by the sensor at a
   293   294   295   296   297   298   299   300   301   302   303