Page 177 - Glucose Monitoring Devices
P. 177

Calibration of minimally invasive CGM sensors  179




                  panel (B), where the IG profile (obtained by convolving a given, simulated, BG
                  profile with a single exponential with s ¼ 11 min) shows both amplitude attenuation
                  and phase delay compared to the BG profile. Notably, s shows inter- and intrasubject
                  variability and its numerical identification requires a suitable collection of both BG
                  and IG samples. Published values of the time constant s range from 6 to 15 min [21].
                  In practice, the BG-to-IG time constant s is treated as a user parameter, but its role in
                  the calibration process needs to be carefully considered [23,24].
                     A second critical aspect behind the differences pointed out in Fig. 9.3 is related to
                  the time variability of sensor sensitivity. The raw electrical current signals acquired
                  by CGM sensors often exhibit a nonphysiological drift, especially on the first day
                  after sensor insertion. An example of nonphysiological drift observed in a raw
                  CGM signal acquired by the Dexcom G4 Platinum CGM sensor is depicted in
                  Fig. 9.5, where the continuous line represents the electrical current signal (in units
                  not specified by the manufacturer) and the dashed line shows the drift. This phenom-
                  enon is related to a variation of sensor sensitivity after its insertion in the body when
                  the sensor membrane enters in contact with the biological environment and un-
                  dergoes the immune system reaction [25,26]. The calibration model has to properly
                  compensate for such time variability, which is often nonlinear.
                     Finally, a third critical aspect is related to the low number of BG samples that are
                  usually available as references for calibration and to the fact that those measures are
                  affected by noise. Indeed, patients usually acquire only a few (e.g., 2e3) SMBG

























                  FIGURE 9.5
                  Representative raw CGM sensor signal (continuous line, units not specified by the
                  manufacturer) that exhibits a nonphysiological drift (dashed line) due to the time
                  variability of sensor sensitivity.
                  Taken from Acciaroli G, Vettoretti M, Facchinetti A, Sparacino G. Calibration of minimally invasive continuous
                          glucose monitoring sensors: state-of-the-art and current perspectives. Biosensors 2018;8(1):24.
   172   173   174   175   176   177   178   179   180   181   182