Page 192 - Glucose Monitoring Devices
P. 192

194    CHAPTER 9 Calibration of CGM systems






























                         FIGURE 9.10
                         Boxplot of performance metrics. From left to right, MARD, PAGE, and percentage of
                         points in A zone of the CEG obtained on the test sets for the original calibration (performed
                         twice per day) and for the Bayesian calibration algorithm with three different calibration
                         frequencies (one calibration per day, one calibration every 2 days and one calibration
                         every 4 days). The stars represent the mean values of each distribution.
                            Adapted from Acciaroli G, Vettoretti M, Facchinetti A, Sparacino G, Cobelli C. Reduction of blood glucose
                         measurements to calibrate subcutaneous glucose sensors: a Bayesian multiday framework. IEEE Transactions on
                                                               Biomedical Engineering 2018;65(3):587e595.
                         per day), 8.94%, 97.06%, 92.65% (Bayesian algorithm, one calibration every
                         2 days), and 10.15%, 93.38%, 87.50% (Bayesian algorithm, one calibration every
                         4 days).
                            The results on the entire dataset are reported via boxplot in Fig. 9.10, where the dis-
                         tributions of the three performance metrics, both for the original calibration and
                         for the Bayesian algorithm (with different calibration frequencies), are shown. In gen-
                         eral, the Bayesian algorithm appears to be more accurate than the original
                         manufacturer calibration, for all the considered metrics and independently from the
                         frequency of calibrations. Indeed, the metrics obtained for the Bayesian calibration
                         algorithm show distributions concentrated at lower MARD values and higher
                         PAGE and CEGA-A values with respect to the original manufacturer calibration. In
                         particular, the Bayesian algorithm with one calibration every 4 days compared to
                         the original manufacturer calibration (on average two calibrations per day) shows
                         11.62% MARD (vs. 12.83%), 89.20% PAGE (vs. 80.62%), and 87.5% CEGA-A
                         (vs. 81%).
                            Numeric values of performance metrics and relative statistical analysis results
                         are reported in Table 9.1. Columns 2e5 report, respectively, the median values of
   187   188   189   190   191   192   193   194   195   196   197