Page 88 - Inorganic Mass Spectrometry : Fundamentals and Applications
P. 88

78                                                         Olesik


            have  a  wide  distribution of velocities. As  a  result,  droplet-droplet  collisions  are
                                      of
            likely, depending  on  the  number droplets  per  unit  volume.  Larger  aerosol  drops
                                                           of
            may collide with  the  walls of the  spray  chamber  as  a  result turbulence  induced
            losses.  Drops  with  sufficient  momentum  may  not  be  able to follow the gas  flow
            through  the  spray  chamber,   so these  drops  impact  on  the  walls   of  the  spray
            chamber  (this is called ine~tial ~e~~s~tiun). Gravitational  settling is another  poten-
            tial  means to lose large droplets  as  they  pass  through  the  spray  chamber.  Evapora-
            tion  results  in  a  decrease  in  aerosol drop size,  thereby  improving the chances for
            the  analyte  to  follow  the gas flow through  the  spray  chamber  and  into the ICP.
                 The spray  chamber is often  erroneously  thought to act as a simple  cut-off
            filter  that  removes  only  droplets  with  diameters  greater  than  some  maximum
            diameter.  However,  even  very  small   (<3-pm  diameter)  droplets  have  a  low
            probability of  passing  through  the spray  chamber  and  reaching  the  plasma  [5]
            when  the  sample  uptake  rate  is 1 dhin (Fig.  3.9).  This  is most  likely  due to
            droplet-droplet  collisions,  coagulation,  and  subsequent  loss of the  larger,  coagu-
                                                                        As
            lated  droplets by turbulence  losses,  inertial  impact,  or  gravitational  settling. is
            also seen  in Fig. 3.9,  droplets  with initial diameters  greater  than  about  15 pm are
            not  transported  through  the  spray  chamber  with  good  efficiency,  even  at  low
            sample  uptake  rates.
                 If the sample  uptake  rate  is decreased,  the  number  of  droplets  per  cubic
            centimeter  in the spray  chamber  decreases,  droplet-droplet  collisions  resulting  in
            coagulation  are  less likely, and  the  analyte  transport  efficiency  increases,  as  shown





                           ,'I,  20 pL/min
                           l
                                                    Gain
                 1 .o                         ""-"""""
                                                    Loss

                 0.5




                 0.0
                     0         5        10       15        20
                              Droplet  Diameter  (vrn)

                     Ratio of  tertiary to primary  aerosol  volume  as a function of  drop size for
            different sample uptake rates. A Getac  microconcentric nebulizer (MCN)  was  used  in  a
                                                                         422.)
            double-pass spray  chamber. Other concentric nebulizers behave  similarly.  (From  Ref.
   83   84   85   86   87   88   89   90   91   92   93