Page 338 - Instrumentation Reference Book 3E
P. 338

Absorption and reflection techniques  321

               The absorption cell is a thick-walled tube with   amplifier is applied to a gain-controlled amplifier
             heavy flanges.  Standard construction  is  in  mild   which produces  an output signal of  3 V peak-to-
             steel, nickel  plated,  but  type  3 16  stainless  steel   peak  mean.  Thus  the  mean  value  of  I, + I,  is
             construction  is  available  where  required.  The   maintained  constant.  The signal is demodulated
             windows  are  of  calcium  fluoride,  sealed  with   and smoothed to obtain the 6 Hz envelope wave-
             Viton O-rings and retaining rings. A heater wire   form.  A  sample-and-hold  circuit  samples  the
             is wound  on the cell, and the sample gas passes   signal level near the end of each half-cycie of the
             through  a  tube  in  thermal  contact  along  the   envelope, and this produces a square wave whose
             length  of  the  cell  before  entering  it  at the end.   amplitude is  related  to  I, - I,,,.  Since Ir + I,  is
             Provision  is  made  for  rodding  out  tubes  and   held  constant,  the  amplitude is  actually propor-
             entries  in  case  of  blockage.  A  thermistor   tional to (Ir  - Im)/(Zr + I,)  which is the required
             embedded in the cell wall detects the cell tempera-   function  to give  a linearized output in terms  of
             ture which is  controlled  by  a circuit in the elec-   sample  concentration.  This  signal  is  amplified
             tronics  unit.  The cell is  thermally  insulated  and   and passed to a phase-sensitive detector, consist-
             sealed  inside  a  plastic  bellows.  The  enclosed   ing of a pair of gating transistors which select the
             space  is  coupled  to  the  purge  system. The two   positive and negative half-cycles and route them
             end units each have a sealing window so these is   to  the  inverting  and  non-inverting  inputs  of  a
             a double seal between the cell and the interior of   differential amplifier. The output of this amplifier
             the detector and source units. Since the source is   provides the 0-5  V output signal.
             inside a further  sealed module, there is minimal   The synchronizing signals for the sample-hold
             danger of the hot source being exposed to leakage   and phase-sensitive detector  circuits are derived
             from the sample cell. The gaps between the three   from  the  torsional  oscillator  drive  circuit  via
             units are normally  sealed with neoprene gaskets,   appropriate time delays. The instrument  span is
             and the whole device is sufficiently well sealed to   governed by selection of feedback resistors in the
             maintain a positive purge pressure of at least 2 cm   low  frequency  amplifier,  and  a  fine  trim  is
             water  gauge  with  a  purge  gas  consumption  of   achieved  by  adjusting  the  signal  level  at  the
             8.3 cm3/s. For use with highly flammable sample   gain-controlled  amplifier. This is a preset adjust-
             gases,  the  sealing  gaskets  at  either  end  of  the   ment-no  operator adjustment of span is consid-
             absorption  cell may  be  replaced  by  vented  gas-   ered  necessary  01-  desirable.  A  front panel  zero
             kets. In this case a relatively large purge flow may   adjustment  is  provided.  This  adds  an  electrical
             be maintained around the cell, escaping to atmos-   offset signal at the phase-sensitive detector. The
             phere across the windows. Thus, any leak at the   system  is  normally  optically  balanced  (Le.;
             windows can be flushed out.              I, = I,)  at  some  specified concentration  of  the
               To facilitate servicing on site, the source, detec-   measured variable (usually zero).
             tor, torsional vibrator,  lamps, pre-amplifier, and   The  current  output  and  alarm  circuits  are
             source .voltage control are all removable without   located  on a  separate printed  circuit board.  The
             the use of a soldering iron. Since the single-beam   voltage output is applied to an operational ampPi-
             system is tolerant to window obscuration and the   fier with  selected feedback  and offset signals to
             internal  walls  of  the  absorption  cell  are  not   produce O-lOmA? 5-20mA,  or 10-50  mA output.
             polished,  cell cleaning will  not  be  required  fre-   The required output is obtained by soldered selec-
             quently,  and  in  many  cases  adequate  cleaning   tor links. The output current is unaffected by load
             may  be  achieved  in  situ  by  passing  solvent  or   resistances  up  to  1 kR  at  50mA;  or  5 kR at
             detergent  through  the  measuring  cell.  There  is   10mA.
             no need to switch the instrument off while doing   A front panel alarm-setting potentiometer pro-
             this. If it becomes necessary the cell can be very   vides a preset signal which is compared with the
             quickly removed and disassembled.        analyzer output voltage in a differential amplifier.
               The electronics unit  contains  the  power  sup-   The output of this opens a relay if the analyzer
             plies together with signal processing circuits, tem-   output  exceeds  a  preset  value,  which  may  be
             perature  control  circuits? output  and  function   either a low or a high analyzer output as required.
             check meter operating controls and signal lamps.   The  alarm  condition  is  indicated  by  two  signal
             The housing is of cast-aluminum alloy, designed   lamps  on  the  panel,  and  the  system  can  be
             for fiush panel mounting. The circuitry is mostly   arranged  to  operate  external  alarms,  or  shut-
             on plug-in printed circuit boards. The indicating   down circuits.
             meter,  controls,  and  signal lamps  are accessible   The power to the cell heater  and the detector
             through a window in the door. The unit is semi-   cooler is controlled from a bridge circuit contain-
             sealed,  and  a  purge  flow  may  be  connected  if   ing thermistors which detect the temperatures of
             sealed  glands  are  used  at  the  cable  entry.  The   the absorption cell and detector.
             signal  processing  circuits  are  contained  on   The indicating meter  on the front panel  has a
             printed  circuit  boards.  Output  from  the  pre-   calibrated output scale, and is used in conjunction
   333   334   335   336   337   338   339   340   341   342   343