346 Introduction to Microfabrication



            efficiency at particle size >0.12 µm. Filter efficiencies  valves, regulators, mass flow controllers, etc.), leak rates
            can also be classified according to most penetrating par-  (static leak test, helium leak test) and gas impurity tests.
            ticle size (MPPS). Filter defects (pinholes) are also a  Bulk gases (also known as line gases or house gases)
            major concern. Air velocity in the cleanroom is usually  are gases shared by many tools. These include nitrogen,
            ca. 0.35 to 0.45 m/s; and air circulation takes place 50  oxygen, hydrogen, argon and compressed air. Nitrogen
            to 500 times/h, depending on cleanliness requirements.  is especially widely used, both in processes and as an
            Once the air has been processed, it is re-circulated, with  inert protective gas. Four purity classes of nitrogen can
            only 10% of replacement air introduced in each cycle.  be offered for different applications:
              Many types of process equipment produce excessive
            heat loads, for example, furnaces in the range of 100 kW,  – process nitrogen: furnace annealing or reactive
            and this heat has to be removed in order to maintain  sputtering, 7N purity;
            constant temperature in the cleanroom. Most of the  – dry nitrogen: venting and flushing of process
            excess heat is taken away by cooling water. The design  chambers, 5N purity;
            of a cleanroom must, therefore, include knowledge of  – pistol nitrogen: for drying;
            the processes and tools that are going to be employed.
                                                         – pump nitrogen: as ballast for pumps.
            35.2.3 DI-water                                Specialty gases are used by dedicated equipment,
                                                         and they are supplied from gas bottles in a one-to-
            De-ionized water (DI-water), also known as ultra-  one distribution topology. These include, for example,
            pure water (UPW), is a major sub-system because of  SF 6 and Cl 2 for etchers, SiH 2 Cl 2 and NH 3 for nitride
            enormous water consumption in modern IC fabrication.  LPCVD, SiH 4 and N 2 O for PECVD oxide, PH 3 for
            A big fab uses a million cubic metres of ultra-pure water  doped polysilicon LPCVD and WF 6 for tungsten CVD.
            a year.                                      Ion implanter gas consumption is very small, and AsH 3 ,
              Water is treated in many steps as follows:  PH 3 and BF 3 mini-bottles are usually located inside the
                                                         implanter cabinet. Implanter gases can also be supplied
             – sand filter;
                                                         from safe delivery system (SDS) sources: the dopant
             – active carbon filter;
                                                         gases are absorbed in solid absorber material in the
             – particle filtering at 3 µm;
                                                         bottle, and released by application of temperature or
             – softening of water;
                                                         underpressure.
             – RO: reverse osmosis;
             – CEDI: continuous electrical de-ionization;
             – UV treatment;                             35.3 ENVIRONMENT, SAFETY AND HEALTH
             – ion exchangers;                           (ESH) ASPECTS
             – particle filtering at 0.2 µm;
             – storage tank;                             Various gases, chemicals and tools are sources of
             – continuous DI-water circulation in the cleanroom  potential health hazards to cleanroom personnel. Ion
               loop.                                     implanters operate at 200 kV and they are sources of
                                                         X-rays (and gamma rays may be emitted in hydrogen
              Reverse osmosis is a process in which water  implantation); plasma systems may leak microwave
            molecules diffuse through a porous membrane, while  energy and UV radiation, and wet etch and plating baths
            microorganisms, particles and ions are rejected. UV  may contain cyanides. These hazards are dealt with in
            treatment kills bacteria and reduces total carbon con-  different ways.
            tent. Both RO and UV treatment can be repeated  Strong mineral acids such as H 2 SO 4 , HNO 3 , H 3 PO 4
            for improved performance. DI-water quality is mon-  and HCl are routinely used. Normal burn hazards are
            itored by resistivity measurements: 18 Mohm-cm is  associated with them and they must be neutralized after
            required. Regular bacteria checks as well as particle tests  use. HF is different because its effect is not immediate
            are performed.                               but delayed, and it does not attack skin but bone. Special
                                                         care is needed for all HF-containing liquids and separate
                                                         disposal of HF is required.
            35.2.4 Gas systems
                                                           Solvents and organics come from various sources:
            Gas system requirements include particle specifications  HMDS, which is used as a priming agent before pho-
            (which set limits to the choice of materials for piping,  toresist coating, is released into cleanroom air (HMDS