Page 313 - Machinery Component Maintenance
P. 313

Balancing  of Machinery Components   295

                   justed by the RSS method. The resulting probable maximum error should
                   ideally not use up more than one tenth of the reciprocal of the specified
                   Unbalance Reduction Ratio factor. For example, if a URR of 95 percent
                   is to be proven, the total test mass error from parameters  1 to 5 should
                   not exceed 0.1 -  5 percent  = 0.5 percent of the test mass weight.
                     Often test masses need to be so small that they become difficult to han-
                   dle. It  is then quite common to work with differential test masses, i.e.
                   two masses  180” opposite each other in the same transverse plane. The
                   effective test mass is the difference between the two masses, called the
                   “differential unbalance.” For instance, if one mass weighs 10 grams and
                   the other 9, the difference of  1 gram represents the differential unbal-
                   ance.
                     When working with differential test masses, the errors of the two com-
                   paratively large masses affect the accuracy of  the differential unbalance
                   in an exaggerated way. In the example used above, each differential test
                   mass would have to be accurate within approximately .025 percent of its
                   own value to keep the maximum possible effect on the differential unbal-
                   ance to within 2 -  0.25 percent  = 0.5 percent. In other words, if the op-
                   posed masses are about ten times as large as their difference, each mass
                   must be ten times more accurate than the accuracy required for the differ-
                   ence.

                                            Test Procedures

                     To test the performance of a balancing machine, IS0  2953 prescribes
                   two separate tests, the U,,,  Test and the Unbalance Reduction Test. The
                   origin and  philosophy behind  these  tests  and  their  purpose  were  ex-
                   plained. Here are the actual test procedures:

                   U,,,   (or Traverse) Test
                     I.  Perform  the mechanical  adjustment, calibration and/or setting of
                        the machine for the particular proving rotor being used for the test,
                        ensuring that the unbalance in the rotor is smaller than five times
                        the claimed minimum achievable residual unbalance for the  ma-
                        chine.
                     2.  Put 10 to 20 times the claimed minimum achievable residual unbal-
                        ance on the rotor by adding two unbalance masses (such as balanc-
                        ing clay). These masses shall not be:
                             in the same transverse plane
                             in  a test plane
                             at the same angle
                             displaced by  180”
   308   309   310   311   312   313   314   315   316   317   318