Page 341 - Machinery Component Maintenance
P. 341

Balancing of  Machinery Components   323








                            \-














              Figure 6-44.  A centrifuge is rebalanced on the test stand in two planes.







       exclusively up  to the present.  They  require experience, accuracy and
       time (approximately 30 minutes). The appearance of relatively inexpen-
       sive programmable pocket calculators (Figure 6-46) makes it possible to-
       day to replace these methods with more accurate numerical methods. The
       determination of amount and angular position of the correction masses
       for both correction planes can be carried out in approximately two-three
       minutes even by untrained personnel.  In detail, the following method
       should be followed:

          1.  Using the balancing instrument, the angular position and amount of
            the unbalance vibration is measured at bearing positions 1 and 2 and
            the values  are entered into the schedule  “initial unbalance  run”
            (Figure 6-47).
         2. The centrifuge is brought to a standstill and a known calibrating
            mass is applied in correction plane A 1. After again reaching the op-
            erational speed the unbalance vibration is measured again and en-
            tered “test run  1.”
         3.  The calibrating mass  is removed from plane A1  and applied to
            plane A2. The resulting  measured values  are again  noted  down
            “test run 2.”
         4.  The evaluation of  the measurement results listed in Figure 6-47
            using the pocket calculator gives the correction masses that must be
   336   337   338   339   340   341   342   343   344   345   346