Page 358 - Microsensors, MEMS and Smart Devices - Gardner Varadhan and Awadelkarim
P. 358

338    IDT MICROSENSOR  PARAMETER MEASUREMENT

                                                    1
  not  so  surprising,  as  acoustic  sensors  are  digital  by  nature  with a  good  dynamic  range
  and  linearity.  Therefore,  state-of-the-art  system  instrumentation can  readily  exploit  the
  digital  acoustic  microsensor.  One  of  the  main  advantages  digital  instrumentation  has
  over  the  earlier  system  technologies  is  that  there  is  no  requirement  to  perform  any
  frequency  to  voltage  (for  output) conversion -  a  process  that  often  causes  a  consider-
  able  loss  of  resolution.  In  the  state-of-the-art  systems,  frequency  to  digital  conversion
  takes  place  without loss 2  through  the  use  of  simple  electronic  counters.  However,  other
  measuring  system  instrumentation  must  be  considered  and  compared  in  terms  of  its
  viability,  functionality, and suitability.  Having accepted that either frequency  or  multiple-
  period measurements (timing signal) is the most appropriate parameter to measure, consid-
  eration  will now  be  given to  past  and  current forms of  such  instrumentation (Campbell
  1998).
     There  are three  electronic  configurations that may be  used  to  measure  the response  of
  a  SAW (two-port)  microsensor:
  1.  Where  the  microsensor  is  connected  to  a  network analyser  (or  vector  voltmeter)  and
     scanned  by  a  narrow  radio  frequency  (RF)  band  on  either  side  of  its  fundamental
     resonant  frequency  (see  Section  11.3).

  2.  Where  the  microsensor  is  used  as  the passive  element,  being  driven  from  a fixed RF
     source  (amplitude or phase  measurements)  (see  Sections  11.4 and  11.5).
  3.  Where the microsensor  is used as the active feedback-determining element, controlling
     the  frequency  of  an oscillator  circuit (frequency  measurement) (see  Section  11.6).


  11.3  NETWORK ANALYSER           AND VECTOR
        VOLTMETER

  Primarily,  the  network  analyser  and  vector  voltmeters  are  used  by  radio  and  electronic
  engineers  for  network  analysis  and  design.  Although  SAW  resonators  have  equivalent
  electric circuit  analogues,  network analysis is usually performed  on  the  basis  of  the  idea
  of  electrically  matching  resonators  with  then-  corresponding  oscillator  circuits.  This  is
  particularly true for the  SAW-sensor oscillator  design.  From either the network  analysers
  or the vector voltmeters, the characteristic  admittance  Y or impedance  Z can be measured.
  These measurements will produce a locus of the admittance or impedance parameters over
  a set of discrete  frequency  data points. The admittance (or impedance)  can be expressed  in
  either rectangular (real and imaginary) or polar  (magnitude and phase) coordinates. Anal-
  ysis  from  such instruments will  deliver  the  most  information  possible  about  an  acoustic
  transducer or the  sensor  that is under load, detailing shifts  in resonant frequency  and any
  changes  in the  quality factor  Q  (Subramanian  1998;  Piscotty  1998).
    Despite  the  advantages  that  either  the  network analyser  (or  vector  voltmeter)  has  in
  delivering  very  detailed  information  on  a  number of  important  parameters,  both  these
  instruments  are  impractical  as  general-purpose  laboratory  tools.  The  reasons  for  this  are
  that  the resonator setup is:

  1
    Digital  in  the  sense that  they  are  active  frequency  elements  in  oscillator  circuits  where  frequency  counters
  are  used  to  measure the output  frequency.
  2
   High  resolutions are achieved  through  long counting periods.
   353   354   355   356   357   358   359   360   361   362   363