Page 130 - Tandem Techniques
P. 130

Page 112

            indicates that the early peaks are the most narrow, and any dispersion that takes place in the connecting
            tube will have the maximum deleterious effect on peaks eluted at k'< 1. It is also seen that the increase
            in peak width with (k') has become fairly small, even when k'=2. It follows, that if the chromatographic
            conditions can be adjusted to ensure that the peaks of interest are all eluted at a (k') of two or more, then
            the detrimental effect of extra column dispersion will be minimized. It is also seen that the optimum
            linear velocity has an unique value for each (k') value of the solute. It follows that the choice of flow
            rate will be optimum for only those solutes that are eluted at and around a specific (k') value. It is clear
            that the optimum velocity should be chosen at the (k') value of the closest eluted solutes where
            separation is the most difficult.


            According to Klinkenberg [4], the variance due to extra column dispersion,       should be restrained

            to be no greater than 10% of the peak variance        to ensure its effect on column resolution is not
            serious. That is,










            It follows that the dispersion from the connecting tube must not be more than one third of the dispersion
            produced by the column. The dispersion from tubes 0.025 cm I.D. (ca .010 I.D.), and 0.0125 cm I.D.
            (ca 0.005 I.D.), when used in conjunction with GC columns 30 m long and 0.025 cm I.D., and 0.053 cm
            I.D., was calculated for the dead volume peak, i.e. the worst case scenario. The results are shown in
            Figure 3.2.

            It is seen that for the GC columns the contribution of the connecting tube to the column dispersion is
            negligible, and far below the advised limit of one third the column dispersion. It is also seen that the
            small diameter tubing need not be used and at least 50 cm of the 0.025 cm I.D. tubing can be tolerated,
            without the inherent dispersion having any serious effect on the column resolution.
   125   126   127   128   129   130   131   132   133   134   135