Page 253 - Vogel's TEXTBOOK OF QUANTITATIVE CHEMICAL ANALYSIS
P. 253

EilUlPMENT FOR HPLC   8.3

       (b) A linear response. The linear range of a detector is the concentration range
           over  which  its  response  is  directly  proportional  to  the  concentration  of
           solute.  Quantitative  analysis  is more  difficult  outside  the linear  range  of
           concentration.
       (c)  Type of response, i.e. whether the detector is universal or selective. A universal
           detector will  sense al1 the constituents of  the sample,  whereas a  selective
           one will only respond to certain components. Although the response of the
           detector will  not be independent  of  the operating conditions, e.g. column
           temperature or flow rate, it is advantageous if the response does not change
           too much when  there are small changes of  these conditions.
          A summary of  these characteristics for different types  of  detectors is given
       in Table 8.2.
       Table 8.2  Typical detector characteristics in HPLC
                                                 -
                         Response   Ca (g mL - ' )  Linear range*
       Amperometric      Selective   10  - l0   104-IO5
       Conductimetric    Selective   10 - '   IO3-104
       Fluorescence      Selective   10- l2   103-104
       UVIvisible absorption   Selective   104-105
       Refractive index   Universal        103-104
       *The range over which the response is essentially linear is expressed
       as  the  factor  by  which  the  lowest  concentration  (C,)  must  be
       multiplied to obtain the highest concentration.
         A detailed description of the various detectors available for use in HPLC is
       beyond the scope of the present text and the reader is recommended to consult
       the monograph by        A brief  account of  the principal  types of  detectors
       is given below.
       Refractive  index  detectors;  These  bulk  property  detectors  are  based  on  the
       change of  refractive index of  the eluant from the column with respect to pure
       mobile  phase.  Although  they  are  widely  used,  the  refractive  index  detectors
       suffer from several disadvantages - lack of  high sensitivity, lack of suitability
       for gradient elution, and the need  for strict temperature  control ( I0.001 OC)
       to operate at their highest sensitivity. A pulseless pump, or a reciprocating pump
       equipped with  a  pulse dampener, must  also be  employed. The effect  of  these
       limitations may to some extent be overcome by  the use of  differential systems
       in which  the column eluant is compared  with a reference flow of  pure mobile
       phase. The two chief types of  RI detector are as follows.
       1. The deflection refractometer (Fig. 8.4), which  measures the deflection  of  a
         beam of monochromatic light by a double prism in which the reference and
         sample cells are separated by a diagonal glass divide. When both cells contain
         solvent of  the same composition, no deflection of  the light beam occurs; if,
         however,  the composition  of  the column mobile phase is changed  because
         of the presence of a solute, then the altered refractive index causes the beam
         to  be  deflected.  The  magnitude  of  this  deflection  is  dependent  on  the
         concentration of  the solute in the mobile phase.
       2.  The  Fresnel  refractometer  which  measures  the change  in  the  fractions  of
         reflected  and  transmitted  light  at  a  glass-liquid  interface  as  the  refractive
   248   249   250   251   252   253   254   255   256   257   258