Page 254 - Vogel's TEXTBOOK OF QUANTITATIVE CHEMICAL ANALYSIS
P. 254

8   COLUMN AN0 THIN4AVER  LlîlUlO CHROMATOCRAPHV

                           Column
                           mobile  phase








       Mirror









                     solvent
       Fig. 8.4  Refractive index detector.

         index of the liquid changes. In this detector both the column mobile phase
         and  a  reference flow of  solvent are passed  through small cells on the  back
         surface of  a prism. When the two liquids are identical there is no difference
         between the two beams  reaching the photocell, but when the mobile phase
         containing solute passes through the ce11 there is a change in the amount of
         light transmitted to the photocell, and a signal is produced. The smaller ce11
         volume (about 3 PL) in this detector makes it more suitable for high-efficiency
         columns  but,  for  sensitive  operation,  the  ce11  windows  must  be  kept
         scrupulously clean.
       Ultvaviolet detectous.  The UV absorption detector is the most widely used in
       HPLC, being  based  on the principle  of  absorption  of  UV  visible light as the
       effluent from the column is passed through a small flow ce11 held in the radiation
       beam.  It  is  characterised  by  high  sensitivity  (detection  limit  of  about
       1 x  10-'g  mL-'  for  highly  absorbing  compounds) and,  since  it  is  a  solute
       property detector, it is relatively insensitive to changes of temperature and flow
       rate. The detector is generally suitable for gradient elution work since many of
       the  solvents  used  in  HPLC  do  not  absorb  to  any  significant  extent  at  the
       wavelengths used for monitoring the column effluent. The presence of air bubbles
       in the mobile phase can greatly impair the detector signal, causing spikes  on
       the chromatogram; this effect can be minimised  by degassing the mobile phase
       prior  to  use,  e.g.  by  ultrasonic  vibration.  Both  single  and  double  beam
       (Fig.  8.5)  instruments  are  commercially  available.  Although  the  original
       detectors  were  single-  or dual-wavelength  instruments (254 and/or  280 nm),
       some  manufacturers  now  supply  variable-wavelength  detectors covering  the
       range 210-800  nm so that more selective detection is possible.
         No account of UV detectors would be complete without mention of the diode
       array (multichannel) detector, in which polychromatic  light is passed  through
       the flow cell. The emerging radiation is diffracted  by  a  grating and then  falls
       on to an array of photodiodes, each photodiode receiving a different narrow-
       wavelength  band.  A microprocessor  scans the  array of  diodes many  times  a
   249   250   251   252   253   254   255   256   257   258   259