Page 241 - Failure Analysis Case Studies II
P. 241

226
                                          ,-   Impulsc Linc






















































               Fig. 1. A general view of the discharge header: (a) shows three pressure taps, two are connected, one is flanged-
               off. (b) shows a close-up of the impulse line, depicting the connection-type, coil, and support.




                                      2.  FAILURE  ANALYSIS
            The failed line fractured at the root of the mounting collar, at the instrument side (see Fig.  1).
           Visual examination showed that the fractured piece was jammed inside the nut (Fig. 3). It was noted
           that the broken  piece was excessively off  centred, indicating possible misalignment of  the tubing
          during  original mounting. A  sample cut from  the failed line was  subjected to chemical analysis
          which confirmed that the tubing material was stainless steel type 3 16, as originally specified.
            Samples of the fracture piece were examined using SEM (Scanning Electron Microscopy). The
          examination showed that the mode of failure was predominantly  fatigue. Figures 4-7  show SEM
           photographs. Figure 4 shows the fracture surface, taken close to the outer layers of the tubing. A
   236   237   238   239   240   241   242   243   244   245   246