Page 303 - Instrumentation Reference Book 3E
P. 303

Measurement techniques: radiation thermometers 287

                                                      Figure  14.55(c) shows the graph against  time of
                                                      the chopped radiation together with the resulting
                                                      electrical signal.

                                                      14.62.3  Optical (~isisnppearing Silrrinent)
                                                      fhemoineter
                                                      Optical radiation  thermometers provide a simple
                                                      and accurate means for measuring  temperatures
                                                      in the range 600 "C to 3000 "C. Since their opera-
                                                      tion requires the eye and judgment  of  an opera-
             Figure 14,53 Mechanism of pyroelectriceffect
                                                      tor, they are not suitable for recording or control
                                                      purposes. However, they provide an effective way
                 6V = SQ/C = RASTIC          (14.40)   of making spot measurements and for calibration
                                                      of total radiation thermometers.
             where C is the electrical capacitance between the   In  construction  an  optical  radiation  therm-
             electrodes.  The  pyroelectric  coefficient  R is  a   ometer is similar to a telescope. However, a tung-
             function of temperature reducing with a non-lin-   sten filament  lamp is placed  at the focus of  the
             ear  characteristic  to  zero  at the  Curie tempera-   objective  lens.  Figure  13 56  shows  the  optical
             ture.                                    arrangement  of  an  optical  radiation  therm-
               When used as a detector in a radiation therm-   ometer.  To  use  the instrument  the  point  where
             ometer, radiation  absorbed at the surface of  the   the temperature is required to be known is viewed
             pyroelectric  slice causes  the  temperature  of  the   through the instrument. The current through the
             detector to rise to a new higher level. At the start   lamp  filament  is  adjusted  so  that  the  filament
             the  charge  on  the  electrodes  will  have  leaked   disappears  in  the  image.  Figure  14.57  shows
             away  through  the  external  electrical  circuit  so   how  the  filament  looks  in  the  eyepiece  against
             there  will1  have  been  zero  voltage  between  the   the background  of the object, furnace, or what-
             electrodes.  As  the  slice  heats  up  a  voltage  is   ever is to have its temperature  measured.  At  (a)
             detected  between  the  two  electrodes.  When  the   the current through the filament is too high and it
             device reaches  its  new  temperature,  losing heat   looks bright against the light from the furnace, at
             to its environment at the same rate as it is receiv-   (c) the current is too IOW while at (b) the filament
             ing  heat  by  radiation,  the  generation  of  excess   is  at  the  same  temperature  as the  background.
             charge on the electrodes ceases, the charge slowly   The temperature  of  the filament is known  from
             leaks awa.y through the electrical circuit, and the   its  electrical resistance.  Temperature  readout  is
             detected  voltage  returns  to  zero.  The  device   achieved either by a meter measuring the current
             detects the change of incident radiation. To detect   through  the filament  or by  temperature  calibra-
             a  constant  flux  of  radiation,  Le..  to measure  a   tions on the control resistor regulating the current
             constant  temperature,  it  is  necessary  to  "chop"   through  the  lamp.  The  filter  in  the  eyepiece
             the incident radiation with a rotating or oscillat-   shown  in  Figure  14.56 passes  light  at  a  wave-
             ing shutter.                             length around 0.65 pm.
               The  physical  construction  of  a  pyroelectric   Lamps for  optical  thermometers are not nor-
             radiation thermometer is essentially identical to a   mally  operated  at  temperatures  niuch  in  excess
             total radiation instrument except for the location   of  1500°C. To  extend  the  range  of  the  instru-
             of the radiation-chopping shutter just in front of   ment beyond  this temperature  a neutral filter of
             the detector. Figure  14.55(aj shows the location   known transmission  factor  can be  placed  in the
             and Figure 14.55(b) a typical profile of the optical   light  path  before  the  lamp.  The  measurement
             chopper in a pyroelectric radiation thermometer.   accuracy  of  an  optical  thermometer  is  typically


                Area of electrode

                                                          High
                                                          impedance
                                                          voltmeter




             Figure 14.54  Pyroelectric detector.
   298   299   300   301   302   303   304   305   306   307   308