Page 307 - Instrumentation Reference Book 3E
P. 307

Measurement techniques: radiation thermometers  291

               7.0 -








             e
             I-   0.4 -

               0.2 -


                                                                      I
                       2           3          4           5           6
                                     Wavelength, pm
             Figure 14.59  Transmittanceof one millimeterof soda-limeglass.
             between  the  target  substance and  the  radiation   temperature resulting in a very non-linear scale.
             thermometer. The normal atmosphere always con-   Second, the  radiation  detectors are  themselves
             tains a small but definite amount of carbon dioxide   sensitive  to  ambient  temperature. This requires
             and  a variable amount of water vapor.  Carbon   either that the radiation thermometer be  main-
             dioxide strongly absorbs radiation  between  4.2   tained at a constant temperature or alternatively
             and 4.4 ,urn and the water vapor absorbs strongly   an ambient temperature sensor is mounted beside
             between 5.6 and 8.0pm and also somewhat in the   the radiation sensor to provide a signal for tem-
             regions  2.6  to  2.9pm;  see  Figure  14.60.  It  is   perature correction.
             obvious  that  these  spectral  regions  should  be   To compensate for these two deficiencies in the
             avoided, particularly in the region of  the water   signal suitable electronic circuits must be used to
             bands. If this is not done the temperature calibra-   provide linearization of the signal and to provide
             tion will vary with path length and also humidity.   automatic temperature correction. It is also neces-
             If the air temperature is comparable to or higher   sary to provide correction for the emissivity of the
             than  the  target  temperature  the  improperly   target.  Typically  the  instrument  itself  carries  a
             designed  infrared  thermometer  could  provide   small “head amplifier” to bring the signal up to a
             temperature  measurements  strongly  influenced   suitable  level  for  transmission  to  the  readout
             by air temperatures.                     instrument. This head amplifier also provides the
                                                      required ambient temperature compensation cir-
                                                      cuits.  The  linearization  and  compensation  for
             14.6.2.6  Signal conditioning for radiation   emissivity are provided at the readout module.
             thermometers
                                                        Some modern instruments provide the whole sig-
             AIthough the output of a radiation thermometer   nal conditioning circuitry in the main instrument
             can be used directly in a voltage or current mea-   itself. Figure 14.61 shows such an instrument. In
             suring instrument this is  unsatisfactory for two   this equipment the output is a 4- to 20-milliamp
             pike reasons. First, the energy radiated by a hot   signal linear  with  temperzture and  compensated
             body is a function of the fourth power of absolute   for ambient temperature.










              2       4       6       8       10      12       14      16
                                       Wavelength,  pm
             Figure 14.60  Atmosphericabsorption of infrared radiation.
   302   303   304   305   306   307   308   309   310   311   312