Page 311 - Instrumentation Reference Book 3E
P. 311

Temperature measurement considerations 295
                          ?



















                                                          t                       t









            Figure 14.62  Problems associated with temperature
            measurement in a stirred vessel.
            Probes  in pipes  or  ducts  There  is  frequently  a
            requirement to measure the temperature of a fluid
            flowing in a pipe. This is usually straightforward,
            but there are still points to watch out for. Figure
            14.63  shows  three  possible  configurations  for
            insertion  into  a  pipe.  The  most  satisfactory
            arrangement  is  to  insert  the  thermometer  probe   I
            into the pipe at a bend or elbow. Figure  14.63(a)   Figure 14.63  Problems associated with location  of
            shows this arrangement. Points to note are:   thermometer probe in pipe: (a) preferred arrangement,
                                                      (b) p!obe  obstructing pipe, (c) sensitive area of  probe not
            (a)  To  ensure  that  the  probe  is  inserted  far   fully immersed, (d) alternative preferred arrangement,
                                                      sensitive portion of probe shaded.
            enough  for  the  sensitive  length  to  be  wholly
            immersed and far enough into the fluid to minim-   When measuring temperature  in large pipes or
            ize thermal conduction from the sealing coupling   ducts it must be remembered that the temperature
            to the sensor.                            profile across the pipe may not be constant. This is
            (S) To insert the probe into the direction of flow   especially true for large flue stacks and air-cond-
            as indicated. The reasons for this are to keep the   itioning ducts. The center liquid or gas is usually
            sensor ahead of the turbulence at the bend, which   hotter (or colder in refrigerated systems) than that
            could cause an error due to local heating, and to   at the duct wall. In horizontal ducts carrying slow-
            remove the effects of cavitation that could occur at   moving air or gas the gas at the top of the duct will
            the tip of a trailing probe.  Figure  14.63(b) shows   be significantly hotter than that at Lhe  bottom of
            the problem that can arise in small pipes where the   the duct. In these circumstances careful considera-
            probe can cause serious obstruction to the flow.   tion  must  be  given  as  to  how  a  representative
              Where it is not possible to put the thermometer   measurement  can  be  obtained;  it  may  well  be
            at a bend  in the pipe it  can be  inserted  radially   necessary  to  make  several  measurements  across
            provided  the  pipe  is  big  enough.  Great  care   the duct and average the readings.
            should  be  taken  to  ensure  complete  immersion
            of  the  sensitive  portion  of  the  probe.  Figure
            14.63(c) illustrates this problem. A better solution   14.7.2.2  Radiation errors
            is diagonal1 insertion  as shown at (d). Again  the   Gas temperature measurements present extra pro-
            probe should point into the direction of flow.   blems compared with temperature measurements
   306   307   308   309   310   311   312   313   314   315   316