Page 309 - Instrumentation Reference Book 3E
P. 309

Temperature measurement considerations 293

            ruled out. They may also not be usable if there is   of analog readouts, since they are not limited to a
            likely to be big ambient temperature variation at   rotary dial.
            the readout point or along the length of the capil-
            lary.
              The next question is: what is the height difference   14.7.1.3  Temperature traiismitters
            between the  thermometer  bulb  and the  readout   On large installations, or where a wide variety of
            position? Long vertical runs using liquid-in-metal   different  measurements  are  being  made  with  a
            thermometers can cause measurement offsets due   wide range  of  instrumentation,  it  is more usual
            to the liquid head in the vertical capillary adding   to transfer the signal from the measurement point
            to  (or the  subtracting from)  the pressure  at  the   to  the  control  area  by  means  of  temperature
            instrument  Bourdon  tube.  In the case of  height   transmitters.  This  has  the  great  advantage  of
            differences greater  than.  say,  10 meters,  liquid   allowing  standardization  of  the  readout  equip-
            thermometers  are  liitely  to  be  unsuitable.  This   ment. Also, in the case of electrical transmission,
            then reduces the choice to gas-filled instruments.   by, say, a 4-20  milliamp signal, the measurement
            A  further  consideration  when  specifying instru-   is much less liable to degradation from electrical
            mentation on a new plant is that it is convenient   interference. Also, the use  of  temperature trans-
            from  itinerary  considerations  to  use  as  many   mitters  allows the  choice of  measurement  tech-
            instruments  of  the  same  type  as  possible.  The   nique to be unencumbered  by  considerations  of
            choice of  instrument is then dictated by the most   length of run to the readout location.
            stringent requirement.                      The  choice  of  electrical or  pneumatic  trans-
              On  large  installations  where  many  different   mission is usually dictated by overall plant policy
            types of instrument are being installed and espe-   rather  than  the needs of  the particular  measure-
            cialip where pneumatic  instrumentation  is used,   ment, in this case temperature.  However, where
            capillary instruments can run into an unexpected   the requirement is for electrical temperature mea-
            psychological hazard.  Not  infrequently  a  hard-   surement for accuracy or other considerations the
            pressed instrument  technician has, on finding he   transmission  will also need to be  electrical. (See
            has  too  long  a  capillary,  been  known  to  cut  a   Part 4, Chapter 29.)
            length  out  of the  capillary  and  rejoint  the  ends
            with  a  compression  coupling.  The  result  is,  of
            course,  disaster  to  the  thermometer.  Where  on   14.7.1.4  Coniputer-coiyimtible measurements
            installation  the capillary tube is found to be  sig-
            nificantly  too  long  it  must  be  coiled  neatly  in   With  the  increasing use  of  computer control  of
            some  suitable  place.  The  choice  of  that  place   plants there is a requirement for measurements to
            may  depend  011  the  type  of  instrument.  In gas-   be compatible. The tendency here is to use thermo-
            filled instruments the location of the spare coil is   couples,  resistance  thermometers,  or  where  the
            irrelevant  but  especially  with  vapor  pressure   accuracy does not need to be so high, thermistors
            instruments it wants to be in a position where it   as the measuring techniques. The analog signal is
            will receive the minimum of ambient temperature   either transmitted to an interface unit at the con-
            excursions to avoid introduction of measurement   trol room or to interface units local to the mea-
            errors.                                   surement.  The  latter  usually  provides  for  less
              FOP installations  with long distances between   degradation of the signal.
             the point of measurement and the control room it   As  most  industrial temperature measurements
             1s  almost  essential to use  an electrical measme-   do not require an accuracy much in excess of 0.5
            ment technique. For long runs resistance therm-   percent  it  is  usually  adequate  for  the  interface
             ometers are to be preferred to thermocouples for   unit to work at eight-bit precision. Higher preci-
             two  principal  reasons.  First.  the  copper  cables   sion  would  normally  only  be  required  in  very
             used for connecting resistance bulbs to their read-   special circumstances.
             out equipment are very much less expensive than
             thermoco qle wire  or  compensating  cable.  Sec-   14.7.1.5  Temperature controllers
             ond, the resistance thermometer signal is a higher
             level  and  lower  impedance  than  most  thermo-   While  thermometers,  in  their  widest  sense  of
            couple signals and is therefore less liable to elec-   temperature  measurement  equipment,  are  used
             trical interference.                     for  readout  purposes,  probably  the  majority  of
              An  added  advantage  of  electrical  measure-   temperature measurenients in industrial  applica-
            ments  is  that,  whether  the  readout  is  iocal  or   tions  are for  control purposes.  There  are there-
             remote.  the  control  engineer  is  given  wider   fore  many  forms  of  dedicated  temperature
             options  as to the  kinds  of  readout  available  to   controllers  on the  market.  As  briefly  described
             him. Not only does he have a choice of analog or   in  Section  14.3.5.1,  the  simplest  of  these  is  a
             digital readout but he can also have a wider range   thermostat.
   304   305   306   307   308   309   310   311   312   313   314