Page 473 - Instrumentation Reference Book 3E
P. 473

456  Electrical measurements































            Figure 20.36  (a) Induction wattmeter; (b) electrostatic wattmeter;  (c) thermocouple wattmeter.


            Le., the power dissipated in R.           20.4.3  Connection of wattmeters
              Thus from these two measurements the power
            in the load can be computed.              There  are  two  methods  of  connecting  a  dyna-
                                                      mometer wattmeter  to the measurement  circuit.
              In the compensated thermal wattmeter employ-
            ing matched  thermocouples as  shown in Figure   These are shown in Figures  20.37(a) and (b). In
                                                      the connection shown in Figure 20.37(a) the volt-
            20.36(c) the  value  of  the  resistance R is  chosen   age coil  is  connected  to  the  supply  side  of  the
            such that

                                                                 Current coil
            The output of the wattmeter can then be shown
            to be given by
                                                             Voltage
                                                             coil
                                                          u
                                                                Series
            where k is a constant of the thermocouples.         resistance
              In  the  compensated  thermal  wattmeter  there   Indication = load power +
            are no errors  due to  the  power  taken  by  either   power loss in current coil
            the current or voltage circuits.                      (a)
              Dynamometer  wattmeters are capable of pro-
            viding an accuracy of order 0.25 percent of FSD
            over a frequency range from d.c. to several kHz.   Current coil
            Induction wattmeters are suitable only for use in
            a.c. circuits and require constant  supply voltage
            and frequency for accurate  operation.  The elec-
            trostatic wattmeter is a standards instrument hav-
            ing  no  waveform  errors  and  suitable  for      Series
            measurements  involving low power factors, such    resistance
            as the measurement  of  iron  loss, dielectric loss,   Indication = load power +
            and the power taken by fluorescent tubes. Ther-    power loss in voltage coil
            mocouple  wattmeters  are  capable  of  providing      (b)
            measurements up to 1 MHz with high accuracy.   Figure 20.37  Wattmeter connection.
   468   469   470   471   472   473   474   475   476   477   478