Page 147 - Modern Analytical Chemistry
P. 147

1400-CH05  9/8/99  3:59 PM  Page 130





              130    Modern Analytical Chemistry


                  5 E KEY TERMS


              aliquot  (p. 111)                 multiple-point standardization  (p. 109)  secondary reagent  (p. 107)
              external standard  (p. 109)       normal calibration curve  (p. 109)  single-point standardization  (p. 108)
              internal standard  (p. 116)       primary reagent  (p. 106)         standard deviation about the
              linear regression (p. 118)        reagent grade  (p. 107)             regression  (p. 121)
              matrix matching  (p. 110)         residual error  (p. 118)          total Youden blank  (p. 129)
              method of standard additions  (p. 110)


                  5 F SUMMARY

              In a quantitative analysis, we measure a signal and calculate the  and the use of an internal standard. The most desirable standard-
              amount of analyte using one of the following equations.  ization strategy is an external standardization. The method of
                                                                 standard additions, in which known amounts of analyte are added
                               S meas = kn A + S reag
                                                                 to the sample, is used when the sample’s matrix complicates the
                                                                 analysis. An internal standard, which is a species (not analyte)
                               S meas = kC A + S reag
                                                                 added to all samples and standards, is used when the procedure
              To obtain accurate results we must eliminate determinate errors  does not allow for the reproducible handling of samples and
              affecting the measured signal, S meas , the method’s sensitivity, k,  standards.
              and any signal due to the reagents, S reag .          Standardizations using a single standard are common, but also
                To ensure that S meas is determined accurately, we calibrate  are subject to greater uncertainty. Whenever possible, a multiple-
              the equipment or instrument used to obtain the signal. Balances  point standardization is preferred. The results of a multiple-point
              are calibrated using standard weights. When necessary, we can  standardization are graphed as a calibration curve. A linear regres-
              also correct for the buoyancy of air. Volumetric glassware can  sion analysis can provide an equation for the standardization.
              be calibrated by measuring the mass of water contained or de-  A reagent blank corrects the measured signal for signals due to
              livered and using the density of water to calculate the true vol-  reagents other than the sample that are used in an analysis. The
              ume. Most instruments have calibration standards suggested by  most common reagent blank is prepared by omitting the sample.
              the manufacturer.                                  When a simple reagent blank does not compensate for all constant
                An analytical method is standardized by determining its sensi-  sources of determinate error, other types of blanks, such as the
              tivity. There are several approaches to standardization, including  total Youden blank, can be used.
              the use of external standards, the method of standard addition,


              5 G Suggested EXPERIMENTS

                  Experiments  Calibration—Volumetric glassware (burets, pipets, and  and internal standards are a common feature of many
                          The following exercises and experiments help connect the material in this chapter to the analytical laboratory.


                                                                     Standardization—External standards, standard additions,
                    volumetric flasks) can be calibrated in the manner described
                                                                   quantitative analyses. Suggested experiments using these
                    in Example 5.1. Most instruments have a calibration sample
                                                                   standardization methods are found in later chapters. A good
                    that can be prepared to verify the instrument’s accuracy and
                                                                   project experiment for introducing external standardization,
                    precision. For example, as described in this chapter, a
                    solution of 60.06 ppm K 2Cr 2O 7 in 0.0050 M H 2SO 4 should
                    give an absorbance of 0.640 ± 0.010 at a wavelength of
                                                                   matrix is to explore the effect of pH on the quantitative
                                                                   analysis of an acid–base indicator. Using bromothymol blue
                    350.0 nm when using 0.0050 M H 2SO 4 as a reagent
                                                                   as an example, external standards can be prepared in a pH 9
                    blank. These exercises also provide practice with using  standard additions, and the importance of the sample’s
                    volumetric glassware, weighing samples, and preparing  buffer and used to analyze samples buffered to different pHs
                    solutions.                                     in the range of 6–10. Results can be compared with those
                                                                   obtained using a standard addition.
   142   143   144   145   146   147   148   149   150   151   152