Page 145 - Modern Analytical Chemistry
P. 145

1400-CH05  9/8/99  3:59 PM  Page 128





              128    Modern Analytical Chemistry


                                               5 D Blank Corrections

                                              In discussing ways to standardize a method, we assumed that an appropriate
                                              reagent blank had been used to correct S meas for signals originating from sources
                                              other than the analyte. At that time we did not ask an important question—
                                              “What constitutes an appropriate reagent blank?” Surprisingly, the answer is not
                                              intuitively obvious.
                                                  In one study, 12  analytical chemists were asked to evaluate a data set consist-
                                              ing of a normal calibration curve, three samples of different size but drawn from
                                              the same source, and an analyte-free sample (Table 5.3). At least four different
                                              approaches for correcting the signals were used by the participants: (1) ignore
                                              the correction entirely, which clearly is incorrect; (2) use the y-intercept of the
                                              calibration curve as a calibration blank, CB; (3) use the analyte-free sample as a
                                              reagent blank, RB; and (4) use both the calibration and reagent blanks. Equa-
                                              tions for calculating the concentration of analyte using each approach are shown
                                              in Table 5.4, along with the resulting concentration for the analyte in each of the
                                              three samples.
                                                  That all four methods give a different result for the concentration of analyte
                                              underscores the importance of choosing a proper blank but does not tell us
                                              which of the methods is correct. In fact, the variation within each method for the
                                              reported concentration of analyte indicates that none of these four methods has
                                              adequately corrected for the blank. Since the three samples were drawn from the
                                              same source, they must have the same true concentration of analyte. Since all
                                              four methods predict concentrations of analyte that are dependent on the size of
                                              the sample, we can conclude that none of these blank corrections has accounted
                                              for an underlying constant source of determinate error.
                                                  To correct for all constant method errors, a blank must account for signals
                                              due to the reagents and solvent used in the analysis and any bias due to interac-




                                                      5 3
                                               Table  .    Hypothetical Data Used to Study Procedures
                                                           for Method Blanks
                                                                         Sample
                                                  a                                         b
                                                W s          S stand     Number          W x              S samp
                                               1.6667       0.2500         1            62.4746          0.8000
                                               5.0000       0.5000         2            82.7915          1.0000
                                               8.3333       0.7500         3           103.1085          1.2000
                                               9.5507       0.8413
                                              11.6667       1.0000                    analyte-free c     0.1000
                                              18.1600       1.4870
                                              19.9333       1.6200

                                                             Calibration equation: S stand = 0.0750 ´W s + 0.1250
                                              Source: Modified from Cardone, M. J. Anal. Chem. 1986, 58, 433–438.
                                              a W s = weight of analyte used to prepare standard solution by diluting to a fixed volume, V.
                                              b W x = weight of sample used to prepare sample solution by diluting to a fixed volume, V.
                                              c Analyte-free sample prepared in the same fashion as samples, but without the analyte being present.
   140   141   142   143   144   145   146   147   148   149   150