Page 356 - Biomedical Engineering and Design Handbook Volume 2, Applications
P. 356

334  DIAGNOSTIC EQUIPMENT DESIGN

                       time compared to most human activities, it is fairly long compared to distances covered by photons
                       traveling at the speed of light. Light travels approximately 30 cm/ns so that a 6 ns duration corre-
                       sponds to a distance uncertainty of about 90 cm, which is the approximate detector ring diameter. As
                       a result, the differential distance of the source between detectors has no observable effect on the tim-
                       ing of the coincidence events in conventional PET systems.
                         The arrival time of the annihilation photons is truly simultaneous only when the source is located
                       precisely midway between the two opposed coincidence detectors. If the source is displaced from the
                       midpoint, there will be a corresponding arrival time interval since one annihilation photon will have
                       a shorter distance to travel than the other. As discussed above, this time differential is too small to be
                       useful in conventionally designed PET systems. However, several of the scintillators used in PET
                       tomographs (e.g., LSO, LYSO) are capable of faster response than the 6 to 12 ns timing discussed
                       above. With appropriate electronics, the coincidence timing window has been reduced to 600 ps for
                       these detectors, yielding a source localization uncertainty of 9 cm. 24,25  Even with that reduction,
                       time-of-flight localization cannot be used to directly generate tomographic images, but it can be used
                       to regionally restrict the backprojection operation to areas where the sources are approximately
                       located. In current implementations, the inclusion of time-of-flight information reduces noise in the
                       reconstructed images by a factor of 2. Time-of-flight PET tomographs were actually commercially
                       available for a short time in the 1980s. These systems used BaF detectors which are very fast, but
                                                                     2
                       unfortunately have very low detection efficiency. As a result, these devices did not compete well with
                       the conventional PET tomographs based on BGO. In 2006, a time-of-flight machine based on LYSO
                       detectors was reintroduced and is now commercially available. 26
                         The only criterion for recording a coincidence event is the overlap of output pulses at the coincidence
                       module. True coincidences occur when a source lies on the LOR defined by two detectors. It is possible
                       that events detected at the two coincidence detectors from sources not on the line of response could hap-
                       pen by chance (Fig. 11.12a). As the count rate at each of the singles detectors increases, the likelihood
                       of false coincidences occurring from uncorrelated events increases. These events are called random or
                       accidental coincidences. The random coincidence rate (R) is directly proportional to the width of the
                       coincidence time window (t) and the product of the singles rate at the two detectors (S and S ):
                                                                                   1    2
                                                      R = 2t S S
                                                             1 2
                       It is easy to see that while the true coincidence event rate is linear with the source activity, the ran-
                       dom coincidence rate increases proportional to the square of the activity. Thus, at high count rates,


                       A                                     B





                                                                          Scatter
                                                                           event
                                     True                             True
                                  coincidence                       coincidence
                                    event                             event


                                   Random
                                    event


                       FIGURE 11.12  (a) Random coincidence. A true coincidence event occurs when the annihilation photons from a
                       single decay are simultaneously detected by opposed detectors. When annihilation photons from multiple decays are
                       detected, false information is registered. (b) If one or more of the annihilation photons is scattered, the apparent LOR
                       does not localize the source. Random and scattered coincidences must be subtracted from the acquisition.
   351   352   353   354   355   356   357   358   359   360   361