Page 82 - Biosystems Engineering
P. 82

Biosystems Analysis and Optimization      63

                   The response of the ground speed to the multisine excitation of
               the pump setting is split up in several parts:
                    •  Response at excited frequencies: The frequency response of the
                      process variable on the frequencies directly excited by the
                      multisine is the linear process response on that frequency.
                      Due to the structure of the excitation sequence, contamina-
                      tion of this response by harmonics is very low. Only second-
                      order harmonics or higher can influence the response. The
                      frequency response of the pump setting is also taken into
                      account on the excited frequencies. The pump setting is the
                      control variable to which the multisine is applied. In practice,
                      the measured pump-setting spectrum is not identical to the
                      applied spectrum because of timing and resolution issues.
                      The frequency response on the pump setting is normalized so
                      that it has a unit response on the first excited frequency. For
                      each process variable, the response on the excited frequency
                      is then divided by this normalized input spectrum on that
                      frequency in order to obtain the transfer function from the
                      pump setting to the process variable. The unit of this transfer
                      function is the unit of the gain between the pump setting and
                      the process variable. Because the input spectrum is more or
                      less zero on the nonexcited frequencies, this rescaling can
                      only be performed on the excited frequencies.
                    •  Response at even nonexcited frequencies: The frequency response
                      on these frequencies is most probably caused by the first har-
                      monic of an excited frequency. If this response is clearly larger
                      than the noise response, there are considerable nonlinearities
                      present in the system.
                    •  Response at odd nonexcited frequencies: This response shows the
                      presence of second- or higher-order harmonics.  A clear
                      response on these frequencies indicates large nonlinearities.
                    •  Noise response: Because multiple periods of the multisine are
                      available—in this case, four full periods—the average process
                      response can be calculated. Noise is defined as the variation
                      around this average response (standard deviation per
                      frequency over the four periods). This means that the system
                      is assumed to be constant during excitation.
                   A multisine excitation is not the standard input for the pump
               setting. This means that some effects that are visible with this excita-
               tion may not need to be modeled in the actual controller. After all,
               only the effects that are encountered during normal controller oper-
               ation need to be modeled. Due to nonlinear effects, the response to
               a multisine excitation may differ considerably from the response to
               a step input.
   77   78   79   80   81   82   83   84   85   86   87