Page 431 - Boiler_Operators_Handbook,_Second_Edition
P. 431

416                                                                               Boiler Operator’s Handbook


               gas which always introduced a problem when there was  circuit. The analyzer didn’t work very well while we
               any amount of oxygen in the water. The sampling sys-  were blowing tubes and indicated low oxygen so the air
               tems had to operate at high velocity to reduce the time  flow went a bit high but the analyzer quit failing every
               between analysis and a response to a change in burner  month.
               operation so leaks in the sample piping was always a     Regular failures of the analyzers and drifting of
               concern.                                             the calibration compelled me to provide an air fuel ratio
                    The advent of the zirconium oxide analyzer made  adjustment independent of the oxygen trim control and
               oxygen trim possible on even small commercial boilers  really limit the trim control range so an analyzer failure
               because the analyzer can be mounted in the boiler or  didn’t produce a hazardous situation or a lot of waste.
               stack to achieve fast sampling and analysis. There were  Figure 11-37 shows a schematic of the air flow loop with
               a few made with sampling systems, some integral to the  this configuration. The summer is set to apply a gain of
               analyzer, and I installed a few before the “in-situ” ana-  0.1 to the input so the full range of output of the oxygen
               lyzers came out.                                     trim controller is reduced to a multiplier adjustment of
                    The in-situ zirconium oxide analyzer doesn’t mea-  ±5%. That not only limits the extent the trim controller
               sure the oxygen content of the flue gas. Before you start  can adjust excess air, it also uses the full range of the
               arguing with me you should read on because it really  trim controller output. When the oxygen controller out-
               doesn’t. The analyzer measures the difference between  put is at 50% the multiplier for the fuel air ratio is 100%,
               the oxygen in the flue gas and a reference gas. Often the  basically one, so the air flow signal flows directly to the
               reference gas is the air around the analyzer and, if the  controls without modification.
               boiler casing, ductwork, or stack leaks that reference     That’s where the output of the oxygen trim control-
               can vary in its oxygen content. Many units still use a  ler should be when everything is initially set up, right
               compressed air source as a reference gas and that can be  in the middle of the range so it can act to increase or de-
               complicated by particles or droplets of oil in the com-  crease the excess air. It’s something to look for after your
               pressed air.                                         controls are tuned. If the output is zero the technician
                    To work the zirconium oxide cell (which is a ce-  didn’t leave the control system any way to decrease the
               ramic substrate coated with the metal oxide) must be  excess air. If it was set up in the summer a reduction will
               heated to a temperature around 1500°F. At that tem-  probably be necessary in the winter when the fan starts
               perature any oil in the compressed air will burn and  pumping colder air. On the other hand, if the output
               deplete some of the oxygen in the reference gas. If your  is well above 50% it limits the amount the system can
               analyzer(s) use compressed air I suggest you provide a  increase the excess air. Lacking any reasonable explana-
               separate compressor for them, one of the inexpensive  tion from the technician, the output of the controller
               oil-free compressors that only has to produce air at 10
               psig or so. Besides, it’s a real waste to dump air you
               compressed all the way up to 100 or 150 psig for use as a
               reference gas. You can size the little compressor to match
               your analyzer needs plus a little for calibration, get far
               less expensive air, and it’s oil free.
                    If you fire oil regularly I suggest you incorporate
               a procedure to prevent damage to your analyzer while
               blowing tubes. Steam soot blowers add a considerable
               amount of moisture to the flue gas when they’re oper-
               ating. The problem with that is that steam has a much
               higher specific heat than air and the heater in a zirco-
               nium oxide analyzer has to really put out to push the gas
               temperature up to 1500°F. It’s not the going up that’s the
               problem, it’s when the soot blower shuts off and all of
               a sudden that heat isn’t needed; the analyzer overheats
               and parts burn out.
                    I  solved  a  problem  with  repeated  failures  of  an
               early model of zirconium oxide analyzer by inserting a
               soot blower header pressure switch in the heater power   Figure 11-37. Air flow loop with limited oxygen trim
   426   427   428   429   430   431   432   433   434   435   436