Design evolution of a flexible robotic bending end-effector for transluminal explorations 301















                                                   Figure 13.14
                           The dimensions for the main conduit module and relevant illustrations.



                                          Table 13.1: Specifications of BEE.

                Flexible length                          35 mm
                Connector tube shaft length              110 mm
                Overall endoscope length                 145 mm
                Shaft diameter                           6 mm
                Weight of endoscopic component           3.201 g
                Articulation range                       76 degrees up/down and left/right
                Height of adjustable stand                21.1 cm
                                                          13.0 cm
                Field of view (camera)                   85 degrees
                Battery                                  No battery required. Connect to an electrical socket.



               The connector tube shaft was only introduced from Prototype 4.0 onward and generally
               serves as a means by which the flexible tip of BEE is delivered to the nasopharyngeal
               cavity. The connector tube shaft has not changed much in terms of its design and is simply
               a long cylindrical tube. Based on the Olympus rigid borescopes X009-015-000-55C and
               X017-015-000-52C that have a working length of approximately 150 mm Olympus [5],as
               well as the Ecleris rigid laryngoscope LR067000 which has a diameter of 6 mm [6], the
               connector tube shaft was correspondingly designed to be approximately 110 mm in length
                                                               2
               and 6 mm in diameter. Together with the 35 mm long flexible tip, the total working length
               of BEE comes to 145 mm, which is comparable to that of a borescope.

               Additionally, the connector tube shaft must also be connected to the gear rack, as the
               translational motion of the connector tube shaft through the nasal orifices is entirely
               motorized and user-controlled. The nonworking end of the connector tube shaft is thus
               cuboidal and is adjoined to said gear rack via a press fit.