Page 134 - Instrumentation Reference Book 3E
P. 134

Measurement of density using resonant elements  119

            8.4.7  Dlp transmitter with bubble tubes   8.5  Measurement of density
            This  very  simple  system,  illustrated  in  Figure   using radiation
            8.14, involves two open-ended tubes, terminated
            with  ‘V notches.  These  are  immersed  in  the   Density measurements by this method  are based
            liquid with the ‘‘V> notches separated by a known   on the principle that absorption of gamma radia-
            fixed vertical distance H  and purged  with a low   tion increases with increasing specific gravity  of
            but steady flow of air (or inert gas) at a suitable   the  material  measured.  These  are  discussed  in
            pressure.                                 Part 3.
              A  dlp  transmitter  connected  between  these   The principal instrumentation includes: a con-
            tubes,  with  the  higher-pressure  side  associated   stant gamma source, a detector, and an indicating
            with the lower ”V” notch, measures the difference   or recording instrument. Variations in radiation
             Ap in hydrostatic pressure at the two points. This   passing through  a fixed volume of  flowing pro-
            is  equal  to  the  density x the  vertical  distance   cess liquid are converted into a proportional elec-
            between the two “V” notches:              trical signal by the detector.
                density = Ap/H
                                                      8.6  Measurement of densit
             Although  this method  is  very  simple and effec-   using resonant elements
             tive,  it  is  unsuitable  for  closed  vessels  or  for
             liquids that may crystallize or involve precipita-   Several density-measuring instruments are based
             tion which might  block the bubble tubes and so   on the measurement of the resonant frequency of
             give rise to erroneous results.          an  oscillating  system such  as a  tube  filled with
                                                      the  fluid  under  test  or  a  cylinder  completely
             8.4.8  Other process considerations      immersed in  the medium. Examples of  each  are
                                                      described in the succeeding sections.
             Agitation in a process tank where density meas-
             urement  is  made  must  be  sufficient  to  ensure
             uniformity  of the liquid. But the velocity of fluid   8.6.1  Liquid density measurement
             at  the  points  where  head  pressure  is  measured   The  Solartron 7835 liquid  density  transducer  is
             must  be  sufficiently Bow  to  avoid  a  significant   shown in Figure 8.15. The sensing element com-
             measurement  error.  Locations  of  side-mounted   prises a single smooth-bore tube through which
             transmitters  should  be  sufficiently  high  above   flows the fluid to be measured. The tube is fixed
             the  bottom  of  the tank  to  avoid  errors  due to   at each end into heavy nodal  masses which are
             them  becoming submerged in  the  sediment that   isolated from the outer case by bellows and liga-
             tends to collect there.                  ments.  Located  along  the  tube  are  the  electro-
                                                      magnetic  drive  and  pick-up  coil  assemblies. In
                                                      operation, the amplifier maintains the tube oscil-
                                                      lating at its natural frequency.
                                                        Since the natural frequency of oscillation of the
                                                      tube is a function of the mass per unit length, it
                                                      must  also  be  a  function  of  the  density  of  the
                                                      flowing fluid. It also follows that the tube should
                                                      be  fabricated  from  material  having  a  low  and
                                                      stable coefficient of expansion. If for reasons  of
                                                      corrosion or wear this is not possible. it is impor-
                                                      tant that the temperature is measured and a sui-
                                                      table  correction  applied  to  the  density  value
                                                      determined from the resonant frequency.
              Bubble
              tube                                      Typically, the  tube  vibrates  at  about  1.3 kHz
              air                                     (when filled with water) and with an amplitude of
              SUPPlV                                  about 0.025 mm. Densities up to 3000 kg/m3 can
                                                      be measured with an accuracy of 0.2 kg/m3 and a
                                                      repeatability  of  0.02 kg/m3. This  contrasts  with
                                                      accuracies of  only about  1 percent  of  span that
                                                      can  be  achieved  with  other  methods,  unless
                                                      extreme care is taken.
                                                        The  response  is  continuous  throughout  its
                                                      operating range with no adjustments of  span  or
             Figure 8.14  D/p cell with bubble tubes.   zero.  Recalibration  is  effected by  adjustment  of
   129   130   131   132   133   134   135   136   137   138   139