Page 415 - Instrumentation Reference Book 3E
P. 415

398  Chemical analysis: gas analysis

                          Actuator                  Actuator

                                                   Sample
                          Sample loop         F loop














                      (a) Inject                lbl Backflush
            Figure 18.19  Schematic diagram of sample and backflush valve.

            maintained in column 2 to effect the final separ-   sample from the next stream is  flowing through
            ation  of  the components  of  the sample. By  this   all common lines.
            means  components  of no interest,  such as high-   Prevention of inter-sample contamination from
            boiling  compounds  or  solvents,  can  be  “back-   valve leakage is accomplished by  locating valves
            flushed” to waste before they reach column 2 or   with respect to pressure drops such that any leak-
            the detector, thus preserving the performance  of   age will flow to vent rather than intermix in com-
            the columns and speeding the analytical proced-   mon  lines. A  typical  flow  arrangement  for  gas
            ure.                                     supplies  to  a  chromatograph  for  multi-stream
              Another   related   technique   commonly   application  is  shown  in  Figure  18.20.  This  is
            employed is called “heart cut,” in which the com-   designed  to  ensure  that  the  sample  and  other
            ponent to be  analyzed  is in  the  “center” of  the   supplies  are  delivered  at  the  correct  flow  rate
            sample profile. In this case the light components   and pressure.  A  pressure-relief valve is  fitted  to
            and heavy components  are flushed to vent  and   protect  the  sampling valve from  excessive pres-
            with  only  the  components  of  interest  actually   sure, and shut-off valves are fitted on all services
            being measured by the detector element.   except bottled gas lines.
              The  gas  sample  must  arrive  at  the  sampling   In  some  applications  additional  conditioning
            valve at or only slightly above atmospheric pres-   of the sample is required. Typical of these would
            sure, at a flow rate typically in the range 10-50 ml   be  trace-heating  of  sample  lines  to  maintain  a
            min-’,  and  be  free  from  dust,  oil,  or  abrasive   sample in a gaseous state, vaporization to change
            particles.  The  sampling system may  also  require   a liquid to a gas. and elimination  of stream con-
            filters,  pressure  or  flow  controllers, pumps,  and   taminants by  mechanical or chemical means.
            shut-off valves  for  control  and  processing of  the
            sample stream. All the components of  the system
            must be compatible with the chemical species to be   18.4.2  Carrier gas
            sampled, and must be capable of withstanding the   The carrier gas transports  the components of the
            range of pressures and temperatures expected.   sample over the stationary phase in the chroma-
              Many  applications require analysis  of  two  or   tographic column. The carrier gas must not react
            more process streams with one analyzer. In these   with  the  sample,  and  for  maximum  efficiency
            instances a sample line from each stream is piped   when  using long columns. it  is advantageous to
            to the analyzer, and sample lines are sequentially   use  a  gas  of  low  viscosity. However,  the  most
            switched through solenoid valves to the sampling   important criterion  in  choosing  a  carrier  gas  is
            valve.  When  multi-stream  analysis  is  involved,   often  the  need  to  ensure  compatibility  with  the
            inter-sample contamination  must  be  prevented.   particular detector in use.
            Contamination  of  samples  can  occur  through   The primary  factors determining the choice of
            valve  leakage  and  inadequate  flushing  of  com-   carrier gas are the effect of the gas on component
            mon lines. To ensure adequate flushing, the cap-   resolution  and  detector  sensitivity.  The  carrier
            acity of common lines is kept to a minimum and   gas and type of detector are chosen so that the
            the stream selection valves are timed so that while   eluted  components  generate  large  signals.  For
            the sample from one stream is being analyzed, the   this reason, helium is generally used with thermal
   410   411   412   413   414   415   416   417   418   419   420