Page 416 - Instrumentation Reference Book 3E
P. 416

Process chromatography 399























             Column
             Detector
                                                               E~~~~~~~~                  1
                                                          ~             Explosion proof
                                                       ,  Bulb   ,snubber   fine temp. control
                                                                      ___
                                            indicator   Press. reg.
                                                   0.34 - 4 bar g
                               Analvret base   Analyzer housing     Casting (temp. control)
             Figure 18.280  Flow diagram of  multi-stream chromatograph with thermal conductivitydetector. Courtesy lnvensys
             Foxboro division.
             conductivity cells because of its high thermal con-   ically. Mechanical controllers are either precision
             ductivity. Hydrogen has a higher thermal conduct-   pressure  regulators  which  maintain  a  constant
             ivity  and.  is  less  expensive  than  helium,  but   pressure  upstream  of  the  column  and  detector,
             because  of  precautions  necessary  when  using   or  differential  pressure  regulators,  which  main-
             hydrogen, helium is preferred where suitable.   tain  a  constant  pressure  drop across  a  variable
               Specific properties of a particular carrier gas are   restriction. The principle of operation of one type
             exploited in other types of detectors, for example,   of  electronic flow controller is  shown  in  Figure
             helium in the helium ionization detector. In special   18.21. A proportion of the gas stream is diverted
             instances a  carrier  gas  other  than  that  normally   via a narrow tube, fitted with an electric heating
             associated with a particular  detector may be used   coil as shown. Sensor coils of resistance wire are
             for other reasons. For example, to measure hydro-   wound on the tube upstream  and downstream of
             gen in trace quantities using a thermal conductiv-   the heating  coil. Heat at a constant  rate  is  sup-
             ity  detector,  it  is  necessary  to  use  a  carrier  gas   plied to the heating coil. Cas passing through the
             other than helium because both helium and hydro-   tube is heated by the coil: and some heat is trans-
             gen have high and similar thermal conductivities.   ferred to the downstream sensor. The sensor coils
             Accordingly,  argon  or  nitrogen  is  used  because   are connected in a Wheatstone bridge circuit. The
             eilher has a much lower thermal conductivity than   out-of-balance signal from the bridge, caused by
             hydrogen, resulting in a larger difference in ther-   the  difference in  temperatures  and  hence  resis-
             mal conductivity and greater output.     tance  of  the  upstream  and  downstream  coils,
              The  flow  rate  of  carrier  gas  affects  both  the   depends  on  the  mass-flow  rate  of  gas  through
             retention  time  of  a  compound  in  the  column,   the  tube,  and  on  the  specific  heat  of  the  gas.
             and the shape of the chromatographic peak and   (See  also  Chapter  1,  Section  1.3.)  The  signal,
             hence the amplitude  of  the detector  signal. It  is   suitably  amplified, can  be  used  to  give a  direct
             therefore essential for the flow rate to be readily   readout  of the flow rate of gas through the tube,
             adjustable to constant  known values. The gas is   and can be used to control the flow by feeding the
             usually supplied from bottles, with pressure redu-   signal to open  or close a regulating valve in the
             cing valves to reduce the pressure to a level com-   main gas line downstream  of the sensing device.
             patible  with  the  flow  control  equipment,  and   In cases where the carrier gas flow rate is con-
             sufficient to give the required  flow rate through   trolled  mechanically, a rotameter  is provided  to
             the column and detector.                 indicate  the  flow  rate.  However,  the  best  indi-
              The flow rate of carrier gas may be measured   cation of correct flow is often the analysis record
             and  controlled  either  mechanically  or  electron-   itself, as the retention times of known components
   411   412   413   414   415   416   417   418   419   420   421