Page 421 - Instrumentation Reference Book 3E
P. 421

404  Chemical analysis: gas analysis

                           Reference   Trigger
                           peak     peak      Marker pips at steps 2.3,  5,6,9, and 10
                      Coarse                                        Backflushed
                      auto
                                                               ______
                                                   auto
                      ’   injection
                                 .-Corrected   retention time-   Backflush time
            Ratio-time mode e   I  ,  t   11  ~Component retention time

                                            Component
                                            width time,
                                              1.e..
                                            lntegrmmn
                                         ~
                                            window
            Figure 18.24 Chromatogram, showing logic and switching steps.
              The power unit provides the low voltage regu-   very low retention time (such as air), and is used
            lated supplies for the analyzer and the processor   as the start point for the ratio timing. If a suitable
            and  may  be  mounted up  to  3 meters  from the   component is not consistently present in the pro-
            processor.                               cess sample, one can be injected into the column
              A typical chromatogram of a sample analyzed   at  the  same  time  as  the  sample,  by  using  the
            by the instrument is shown in Figure 18.24, anno-   second loop of the sample valve.
            tated  to  show  the  various  switching  and  logic
            steps during the analysis.                Trigger peak  The trigger peak  must appear on
              The  operation  of  the  chromatograph can  be   the  chromatogram  between  the  reference  and
            either  on  a  fixed-time  or  ratio-time  basis.  In   component peaks. It must be self-evident by vir-
            fixed-time operation  the  sample  injection  is  the   tue of size, and it must be consistent in height and
            start of the time cycle. At preset times the “inte-   width. As with the reference peak it can be from a
            gration window” is opened and closed, to coincide   component  of  the  process  sample,  or  injected
            with the start and finish of the emergence of the   separately. Alternatively it can be a negative peak
            components from the column. While the window   derived  by  using a  doped  carrier  gas. The logic
            is open the detector signal is integrated to give a   circuits measure the time between reference and
            measure of  the concentration  of the component.   trigger peaks and use this, together with the pre-
            Other  operations  such as column  switching and   set ratio value, to compute the time for the start
            automatic zeroing  are similarly timed  from the   of  the integration window. Similarly, the trigger
            sample injection. For fixed-time operation  to be   peak  width  is  used  to  define  the  width  of  the
            reliable, pressure and flow rate of carrier gas, and   window. At the start of  integration the value of
            temperature  and quantity of  stationary phase in   the  signal  level  is  stored.  The  integrator  then
            the column must be closely controlled.    measures  the  area  under  the  component  peak
              Many  of  the  problems  associated  with  fixed-   for  the  period  of  the  window  opening.  At  this
            time operation  may be  avoided by  use  of  ratio-   point the signal level is again measured and com-
            time  operation.  In  this  mode  of  operation  the   pared  with  the  stored  start value  to  determine
            retention  time of components  is measured  from   whether any baseline shift has occurred. The inte-
            an early reference peak (corrected retention time;   gration is corrected for any baseline shifts.
            see  Figure  18.24) instead  of  from  the  time  of   The final value of the integration is stored and
            sample  injection.  The  ratio  of  two  corrected   used  to  give an output  signal, which represents
            retention  times  (retention  ratio)  is  less  affected   the concentration  of the component. As this sig-
            by  changes  in  the  critical  column  parameters.   nal  is  updated  after  each  analysis  the  output
            The corrected  retention  time for an early trigger   shows the trend of the concentration.
            peak is used to predict the time of emergence of   After  the  completion  of  integration,  the  col-
            the component of interest, that is, the integration   umn is backflushed in order to remove any later
            window. For the system to be able to operate in   components,  the  duration  of  the  backflushing
             the ratio mode, it is necessary to have two specific   being  ratioed  from  the  analysis  time.  Alterna-
             peaks  in  the  chromatogram in  advance  of  the   tively, for those applications requiring a measure-
             peak of the component of interest.
                                                      ment such as “total heavies,” the peak of the total
                                                      backflushed components can be integrated.
             Reference peak  The reference peak is due to the   There are some applications where the ratio-time
            first component  eluted from the column, with  a   mode cannot be used. Typically, the measurement
   416   417   418   419   420   421   422   423   424   425   426