Page 426 - Instrumentation Reference Book 3E
P. 426

Calibration of gas analyzers  409

                                                      automatic electronic subtraction of the NO con-
                                                      centration from the NO,  value.
                                                        The flow system of  a nitrogen  oxides analyzer
                                                      is shown in Figure 18.31. Ozone is generated from
                                                      ambient air by the action of W light
                                                             hV
                                                          3024203
                                                      and  a  controlled  flow  rate  of  ozonized  air  is
                                                      passed to the reaction chamber for reaction with
                     Check valve
                CH in-         150cctank              NO in the air sample, which is passed through the
                24
                                            Pressure   chamber  at  a  controlled  flow  of  1 lmin-’.  By
                           ia)              gauge
                                                      selection of  a  switch to  operate  the appropriate
                                                      solenoid valves. a span gas may be directed to the
               Buff
                                                      reaction  chamber,  or a zero calibration may be
                                                      carried  out by  shutting off  the flow of  ozonized
                                                      air to the reactor.  The three-way solenoid valve
                                                      downstream  of the converter is switched to per-
                                                      mit  NO analysis when  bypassing  the  converter,
                                                      and  NO,  analysis  when  the  sample  is  passed
                                                      through the converter. The analyzer can measure
                                                      ozone in air in the range 5 ppb to 25 ppm, with a
                                                      precision of fl percent.
                              yeze
               Thermistor            Sohd             18.5.4  Summary of special gas analyzers
               temperature   SUPP~   WlW
               control                                The operating principles of analyzers for the most
                           Ibl                        commonly  measured  gases  are  given  in  Table
             Figure 18.30  Ozone analyzer. Courtesy Columbia
             Scientific Industries Corp.: (a) flow diagram, (b) functional   18.4.
             block diagram.
                                                      18.6  Calibration of gas analyzers
                       es of nitrogen analyzer
                                                      None  of  the  commonly  used  gas  detectors  is
             Analyzers  for  oxides  of  nitrogen-NO,   NO,   absolute; that is, they are devices where the out-
             (total  oxides of  nitrogen),  NO2  - are based  on   put signal from the detector  for the gas mixture
             the  chemiluminescent  reaction  of  nitric  oxide   under test is compared with that for mixtures of
             (NO)  and  ozone  to  produce  nitrogen  dioxide   the bulk gas containing known concentrations of
             (NO2). About 10 percent of the NO2 is produced   the  determinand.  The use  of  standard  gas  mix-
             in an electronically excited state, and undergoes a   tures is analogous to the use of standard solutions
             transition  to the  ground  state,  emitting  light in   in solution chemistry, but  their preparation  and
             the wavelength range 590-2600  nm:       handling  present some peculiar problems. As  in
                 NO+03 4 NO;+02                       solution  chemistry, the calibration gas mixtures
                                                      should reflect, as closely as possible, the compos-
                 NO;  + NO2 + IZV                     ition of the samples they are desired to measure.
             The intensity of the light emission is proportional   Ideally  a  number  of  standard  mixtures.  whose
             to the mass-flow rate of NO through the reaction   concentration  covers the range of  samples to be
             chamber  and  is  measured  by  a  photomultiplier   measured,  should  be  used  to  establish  the
             tube.                                    response  curve  of  the  instrument  or  detector.
               Analysis of  total  oxides of  nitrogen  (NO,)  in   However,  for  routine  calibration  where  the
             the  sample  is  achieved  by  passing  the  gases   response  curve  has  previously  been  established
             through  a  stainless  steel  tube  at  600-800°C.   or is well known. it is usual to calibrate gas ana-
             Under  these  conditions,  most  nitrogen  com-   lyzers by use of a “zero” gas mixture which is free
             pounds (but not N20) are converted to NO which   of the determinand and es’zablishes the zero of the
             is  then  measured  as  above.  Nitrogen  dioxide   instrument, and one or more “span” gases contain-
             (NO21 may  be  measured  directly by  passing the   ing  concentrations  of  the  determined  close  to
             air sample over a molybdenum catalyst to reduce   those it is desired to measure.
             it  to  NO,  which  is  again  measured  as  above,   The accuracy to which  a  gas mixture  can  be
             or  the  NO2 concentration  may  be  obtained  by   prepared  depends on the number and nature of
   421   422   423   424   425   426   427   428   429   430   431