Page 450 - Instrumentation Reference Book 3E
P. 450

Measurement of d.c. and a.c. current and voltage using indicating instruments 433

                     Current-carrying terminals       ment with a FSD of  lOmA it is l00R/V. If FSD
                                     Four-terminal shunt   requires  only  10pA  then  the  resistance  has  a
                                                      value  of  100,00OR/V. Thus  for  a  voltmeter  to
                                                      have  a  high  input  impedance  the  instrument
                                                      movement must have a low current for FSD.


                                                      20.2.1.2  Characteristics ofpermanent magnet-
                                                      moving coil instruments
                                                      Permanent magnet-moving coil instruments have a
                                                      stable calibration, low power consumption, a high
                                                      torque-to-weight ratio, and can provide a long uni-
                                                      form scale. They can have accuracies of up to 0.1
                                                      percent of  FSD. With  the use  of  shunts or  series
                                                      resistors they can cover a wide range ofcurrent and
                                                      voltage. Theerrorsdue to hysteresis effectsare small,
                                                      and they are generally unaffected by stray magnetic
                                                      fields. It is possible to adjust the damping in such
                                                      instruments to any required value. The major errors
                              (b)                     are likely to be caused by  friction in  the bearings
             Figure 20.10  (a) Ammeter usingarnoving-coil   and changes in the resistance of the coil with tempera-
             instrument; (b) voltmeter using a moving-coil instrument   ture. Copper wire, which is used for the coil, has a
                                                      temperature coefficient of +0.4%/K. When used as
            where R, is the resistance of the shunt, R,  is the   a  voltmeter  this  temperature  variation  is  usually
             resistance of the coil and swamping resistance and   swamped by the series resistance. When used as an
             S is the sensitivity of the unshunted movement.   ammeter with manganin shunts it  is  necessary  to
              High-current  ammeters  usually  employ  a   swamp the  coil  resistance with  a  larger  resistor,
            movement requiring  15 mA for FSD. The shunts   usually manganin, as shown in Figure 20.10(a). This
             are usually four-terminal devices made of manga-   has the effect of more closely matching the tempera-
            nin.  The  voltage  drop  across  the  instrument  is   ture coefficient of the coillswamp resistance combin-
             0.075V  and  thus  the  power  dissipated  in  the   ation to that of the shunt, thus effecting a constant
             shunt is approximately 0.075 IW. Table 20.3 gives   current  division between  the  instrument and  the
             the  power  dissipation  in  the  shunt  for  various   shunt over a given temperature range.
            current ratings.
              For use as a d.c. voltmeter the Sensitivity, S,,  is
             given by                                 20.2.1.3  A.c.  voltage ar?d current measurenzent
                                                      using moving-coil instruments
                                                      The direction of  the torque generated in a mov-
                                                      ing-coil instrument is dependent  on  the instant-
            where R, is the  series resistance, R is  the resist-   aneous direction of  the current through  the coil.
             ance  of  the  coil.  and  S  is  the  sensitivity of  the   Thus  an  alternating  current  will  produce  no
            movement (Figure 20.10(b)).               steady-state deflection.
              The  value  of  the  series resistance depends  on   Moving-coil instruments are provided with an
            the sensitivity of  the  moving  coil.  For  a move-   ax. response  by  the  use  of  a  full-wave bridge
                                                      rectifier,  as  shown  in  Figure  20.11.  The  bridge
            Table 20.3  Power dissipated in shunt for various current
             ratings
                                                                       Moving-coil
             Current (A)           Power dissipated  ( W)        /instrument   -\
               1                    0.075
               2                    0.150
               5                    0.375
              10                    0.75
              20                    1.50
              50                    3.15
             100                    7.50
             200                    15.00
             500                   37.50                   Ammeter           Voltmeter
             1000                  75.00              Figure 20.11  A.c. current andvoltage measurement
                                                      using a rectifier-moving-coil instrument.
   445   446   447   448   449   450   451   452   453   454   455