Page 522 - Instrumentation Reference Book 3E
P. 522

Wavelength ana color  505






                                                                  f   Monochromator  1























                                                                          ib)
             Figure 2+l 1.41  Double-beam techniques in s~ec~~o~h~toin~tr~. ~Linear~ty~me~hod; "optical nuli" method.
                                                       (a)
                                                                     (b)
               Spectrophotometers  may  be  either  "single   attenuator, which may take the form of a variable
             beam," in which the light beam takes a single fixed   aperture or a system of polarizing prisms.
             path and the measurements are effected by taking   Since  spectrophotometric  results  are  nearly
             measurements with and without the sample pres-   always used in  extensive calculations  and  much.
             ent, or "double beam," in which the light is made   data can easily be collected, spec~roph~tonie~ers
                                                                           with
             to  pass  through  two  paths,  one  coiitaining the   are  sometimes  equip~e~  inic~opr~cessors.
             sample and the  other a reference; the  intensities   These microprocessors are also commonly used to
             are then compared. In work on chemical solutions   control a variety of automatic functions, for exan-
             the reference beam is usually passed through a cell   ple, the wavelength-scanning mechanism, automatic
             identical to that of the sample containing the same   sample changing, and SO on.
             solvent so that. solvent and cell effects are cancelled   For chemical purposes  it is nearly always the
             out:  in  reflection  work  the  reference  sample  is   absorbance  (i.e.,  optical  density)  of the  sainple
                                                                             that is required:
             usually  a  standard  white  reflector.  The  single-   rather than the ~rans~~ission
             beam  technique  is  usual  in  manually  operated   1
             instruments  and  the  double-beam  one  in  auto-   A = log,,-
             matic instruments, nowadays the majority.            T
               Two  main  varieties  of  double-beam  techn-   where A is the absorbance or optical density and T
             iques are used (Figure 21.11). That shown in Fig-   is the transmission. The  relation between  trans-
             ure 21.14(a> relies for accuracy on the linearity of   mission and absorbance is shown in Table 2 1.1.
             response of the detector  and is sometimes called   Instrumen~s for chemical work usually read in
             the "iinearity method." The light beam is made to   absorbance only. Wave number---the number  of
             follow alternate sample and reference paths and   waves in one centimeter-is  also used by chemists
             the detector  is used  to measure the intensity of   in preference to wavelength. The relation between
             each in turn; the rati'o gives the transmission  or   the two is shown in Table 21 2.
             reflection factor at the wavelength involved. The   It is not economic to build all-purpose spectro-
             other  method  shown  (Figure 21.14(b)) is called   photometers in view of their use in widely differ-
             the '~opt~ca~-~uI~ method."  Here  the intensity of   ent fields. The most comin~n varieties are:
             the reference beam is reduced to equal that of the
             sampie  beam  by  some form  of  servocontrolled   (If  Transinission/absorba3lce, ultraviolet,   and
             optical attenuator, and the detector  is called on   visible range (200-600  nm);
             only to judge for equality between the two beams.   (2)  Transmission  and  reflection,  near-ultravio-
             The  accuracy  thus  depends  on  the  optical   let, and visible (30&S00 nm):
   517   518   519   520   521   522   523   524   525   526   527