Page 72 - Mechanical Engineers' Handbook (Volume 2)
P. 72

6 Measurement Systems  61



























                                    Figure 22 Matching power requirements for a mass load with dissipation.




                           voltage in a circuit, for example, we must draw some current, at least instantaneously, so
                           that the power is dissipated by effective resistances, or energy is stored capacitively, induc-
                           tively, or both.
                              At the measurement interface, we wish to disturb the measured system as little as pos-
                           sible by extracting as little energy or power as can be managed. It is usually our objective
                           to pass power or energy along the chain of elements that form the measurement system, and
                           there is a best combination of the energy variables to attain the optimum transfer. This chapter
                           has dealt with these issues: the maximization of energy transfer within the system where we
                           want it and the minimization of energy theft at the measurement interface where we do not.


            6.1  Interaction in Instrument Systems
                           The generalized instrument consists of a number of interconnected parts with both abstract
                           and physical embodiments. An orifice flow-metering system might consist of the following
                           chain: An orifice plate converts the flow to a pressure drop; a diaphragm converts the pressure
                           drop to a force; a spring (perhaps an unbonded strain-gage bridge) converts the force to a
                           displacement; the strain gages convert the displacement to a resistance change; a bridge
                           arrangement converts the resistance change to a differential voltage; and a galvanometer
                           converts the voltage to a trace on paper. Figure 23 illustrates this chain.
                              In setting up this chain, the orifice is sized to minimize the pressure drop resulting from
                           our flow measurement, and the diaphragm must be sized for minimum pumping volume
                           during transients in pressure, or it will alter the flow reading in a transient sense. The
                           diaphragm, however, has an output stiffness; the force it transmits decreases with increasing
                           displacement, and it is driving the unbonded gages, converting force to displacement. The
                           combined stiffness of the unbonded strain gages, which will be linear springs, must equal
   67   68   69   70   71   72   73   74   75   76   77