Page 150 - MODELING OF ASPHALT CONCRETE
P. 150

128    Cha pte r  F i v e


                    Specimen Fabrication
                    Asphalt mixtures were mixed and compacted at temperatures that were in accordance
                    with the requirements for each binder. All mixtures were aged at 135°C for 4 hours (i.e.,
                    short-term oven aging) before compaction.
                       For mixtures tested in the axial compression mode, samples were compacted into
                    gyratory plugs of 150 mm in diameter by 178 mm in height. Later, they were cut and cored
                    to cylindrical specimens with dimensions of 100 mm in diameter and 150 mm in height. For
                    mixtures tested in the IDT mode, samples were compacted into gyratory plugs of 150 mm
                    in diameter by 60 mm in height, and they were cut to 38 mm height. Both ends were cut to
                    ensure a more consistent air void distribution along the height of the test specimens.
                       The target air voids for the final cut and cored specimen were 4% ± 0.5%. In order
                    to achieve this density, the target air voids for the gyratory plug had to be higher than
                    that of the cut and cored specimen. The difference between the target air voids for the
                    gyratory plug and for the cut and cored specimen typically increased as the NMAS
                    increased with an average being about 1.5% to 2%.
                       For IDT specimens, a small study was conducted using four mixtures with different
                    NMASs to estimate the reduction in air voids when the specimen is cut from a 60 mm
                    height to 38 mm. The study showed that, in general, the reduction in air voids is around
                    1% for 9.5-mm NMAS and the reduction increases by approximately 0.5% increments
                    as the NMAS increases to 12.5, 19, and 25 mm.
                       Air voids were measured using the Corelok vacuum sealing device. Specifications
                    provided in ASTM D6752-03 were followed in taking these measurements and making
                    the calculations. Appropriate adjustments were made to account for the density of
                    water when measurements were taken at temperatures other than 25°C.
                       For each of the 24 mixtures there were 3 replicates of each tested. During fabrication
                    or testing, if errors were made or densities were not met, then the specimen was
                    discarded and an additional specimen was manufactured and tested.

                    Testing and Analysis

                    Test Setup
                    Testing was performed using a closed-loop servo-hydraulic machine, manufactured by
                    Material Testing System (MTS). A temperature chamber, cooled by liquid nitrogen, was
                    used to control the test temperature. Dummy specimens with thermocouples embedded
                    in the middle of the specimen were used to monitor the temperatures to which the
                    specimens were subjected.
                       For axial compression testing, 100-mm-diameter metal end plates attached to the top
                    and bottom rams were used to apply the compressive load to the specimen. Friction-
                    reducing end treatments were used to lessen the confinement that occurs as a result of
                    the friction against the specimen and the metal end plates. The end treatments were
                    made of two latex membranes, each 0.0125 in. thick. A very small amount of silicon
                    grease was smeared between the latex membranes to allow the specimen to move (dilate)
                    with respect to the metal end plates. Caution was used to apply only the minimum
                    amount of grease necessary because excess grease creates a slick interface that causes the
                    specimen to slip out of the test setup under high load and high frequency combinations.
                    Also, an approximate 25-mm-diameter hole was introduced in the center of the membrane
                    to allow a small amount of contact between the metal plate and the specimen, thereby
                    increasing the friction and reducing the movement of the specimen during testing.
   145   146   147   148   149   150   151   152   153   154   155